JOURNAL ARTICLE

Assessment of intracranial pressure with ultrasonographic retrobulbar optic nerve sheath diameter measurement

Dachuan Liu, Zhen Li, Xuxiang Zhang, Liping Zhao, Jianping Jia, Fei Sun, Yaxing Wang, Daqing Ma, Wenbin Wei
BMC Neurology 2017 September 29, 17 (1): 188
28962603

BACKGROUND: Ultrasonograpic retrobulbar optic nerve sheath diameter (ONSD) measurement is considered to be an alternative noninvasive method to estimate intracranial pressure,but the further validation is urgently needed. The aim of the current study was to investigate the association of the ultrasonographic ONSD and intracranial pressure (ICP) in patients.

METHODS: One hundred and ten patients whose intracranial pressure measured via lumbar puncture were enrolled in the study. Their retrobulbar ONSD with B-scan ultrasound was determined just before lumber puncture. The correlation between the ICP and the body mass index (BMI), ONSD or age was established respectively with the Pearson correlation coefficient analysis. The discriminant analysis was used to obtain a discriminant formula for predicting ICP with the ONSD、BMI、gender and age. Another 20 patients were recruited for further validation the efficiency of this discriminant equation.

RESULTS: The mean ICP was 215.3 ± 81.2 mmH2 O. ONSD was 5.70 ± 0.80 mm in the right eye and 5.80 ± 0.77 mm in the left eye. A significant correlation was found between ICP and BMI (r = 0.554, p < 0.001), the mean ONSD (r = 0.61, P < 0.001), but not with age (r = -0.131, p = 0.174) and gender (r = 0.03, p = 0.753). Using receiver operating characteristic (ROC) curve analysis, the critical value for the risk mean-ONSD was 5.6 mm from the ROC curve, with the sensitivity of 86.2% and specificity of 73.1%. With 200 mmH2 O as the cutoff point for a high or low ICP, stepwise discriminant was applied, the sensitivity and specificity of ONSD predicting ICP was 84.5%-85.7% and 86.5%-92.3%.

CONCLUSIONS: Ophthalmic ultrasound measurement of ONSD may be a good surrogate of invasive ICP measurement. This non-invasive method may be an alternative approach to predict the ICP value of patients whose ICP measurement via lumbar puncture are in high risk. The discriminant formula, which incorporated the factor of BMI, had similar sensitivity and higher specificity than the ROC curve.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
28962603
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"