Efficacy of Subcutaneous Electrocardiogram Leads for Synchronous Timing During Chronic Counterpulsation Therapy.
ASAIO Journal : a Peer-reviewed Journal of the American Society for Artificial Internal Organs 2017 March
Counterpulsation devices (CPDs) require an accurate, reliable electrocardiogram (ECG) waveform for triggering inflation and deflation. Surface electrodes are for short-term use, and transvenous/epicardial leads require invasive implant procedure. A subcutaneous ECG lead configuration was developed as an alternative approach for long-term use with timing mechanical circulatory support (MCS) devices. In this study, efficacy testing was completed by simultaneously recording ECG waveforms from clinical-grade epicardial (control) and subcutaneous (test) leads in chronic ischemic heart failure calves implanted with CPD for up to 30 days. Sensitivity and specificity of CPD triggering by R-wave detection was quantified for each lead configuration. The subcutaneous leads provided 98.9% positive predictive value and 98.9% sensitivity compared to the epicardial ECG leads. Lead migration (n = 1) and fracture (n = 1) were observed in only 2 of 40 implanted leads, without adversely impacting triggering efficacy due to lead redundancy. These findings demonstrate the efficacy of subcutaneous ECG leads for long-term CPD timing and potential use as an alternative method for MCS device timing.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app