SU-E-T-416: Experimental Evaluation of a Commercial GPU-Based Monte Carlo Dose Calculation Algorithm.

Medical Physics 2015 June
PURPOSE: A new commercial GPU-based Monte Carlo dose calculation algorithm (GPUMCD) developed by the vendor Elekta™ to be used in the Monaco Treatment Planning System (TPS) is capable of modeling dose for both a standard linear accelerator and for an Elekta MRI-Linear accelerator (modeling magnetic field effects). We are evaluating this algorithm in two parts: commissioning the algorithm for an Elekta Agility linear accelerator (the focus of this work) and evaluating the algorithm's ability to model magnetic field effects for an MRI-linear accelerator.

METHODS: A beam model was developed in the Monaco TPS (v.5.09.06) using the commissioned beam data for a 6MV Agility linac. A heterogeneous phantom representing tumor-in-lung, lung, bone-in-tissue, and prosthetic was designed/built. Dose calculations in Monaco were done using the current clinical algorithm (XVMC) and the new GPUMCD algorithm (1 mm3 voxel size, 0.5% statistical uncertainty) and in the Pinnacle TPS using the collapsed cone convolution (CCC) algorithm. These were compared with the measured doses using an ionization chamber (A1SL) and Gafchromic EBT3 films for 2×2 cm(2), 5×5 cm(2), and 10×10 cm(2) field sizes.

RESULTS: The calculated central axis percentage depth doses (PDDs) in homogeneous solid water were within 2% compared to measurements for XVMC and GPUMCD. For tumor-in-lung and lung phantoms, doses calculated by all of the algorithms were within the experimental uncertainty of the measurements (±2% in the homogeneous phantom and ±3% for the tumor-in-lung or lung phantoms), except for 2×2 cm(2) field size where only the CCC algorithm differs from film by 5% in the lung region. The analysis for bone-in-tissue and the prosthetic phantoms are ongoing.

CONCLUSION: The new GPUMCD algorithm calculated dose comparable to both the XVMC algorithm and to measurements in both a homogeneous solid water medium and the heterogeneous phantom representing lung or tumor-in-lung for 2×2 cm(2)-10×10 cm(2) field sizes. Funding support was obtained from Elekta.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app