Fully automatic rhythm analysis during chest compression pauses.
Resuscitation 2015 April
AIM: Chest compression artefacts impede a reliable rhythm analysis during cardiopulmonary resuscitation (CPR). These artefacts are not present during ventilations in 30:2 CPR. The aim of this study is to prove that a fully automatic method for rhythm analysis during ventilation pauses in 30:2 CPR is reliable an accurate.
METHODS: For this study 1414min of 30:2 CPR from 135 out-of-hospital cardiac arrest cases were analysed. The data contained 1942 pauses in compressions longer than 3.5s. An automatic pause detector identified the pauses using the transthoracic impedance, and a shock advice algorithm (SAA) diagnosed the rhythm during the detected pauses. The SAA analysed 3-s of the ECG during each pause for an accurate shock/no-shock decision.
RESULTS: The sensitivity and PPV of the pause detector were 93.5% and 97.3%, respectively. The sensitivity and specificity of the SAA in the detected pauses were 93.8% (90% low CI, 90.0%) and 95.9% (90% low CI, 94.7%), respectively. Using the method, shocks would have been advanced in 97% of occasions. For patients in nonshockable rhythms, rhythm reassessment pauses would be avoided in 95.2% (95% CI, 91.6-98.8) of occasions, thus increasing the overall chest compression fraction (CCF).
CONCLUSION: An automatic method could be used to safely analyse the rhythm during ventilation pauses. This would contribute to an early detection of refibrillation, and to increase CCF in patients with nonshockable rhythms.
METHODS: For this study 1414min of 30:2 CPR from 135 out-of-hospital cardiac arrest cases were analysed. The data contained 1942 pauses in compressions longer than 3.5s. An automatic pause detector identified the pauses using the transthoracic impedance, and a shock advice algorithm (SAA) diagnosed the rhythm during the detected pauses. The SAA analysed 3-s of the ECG during each pause for an accurate shock/no-shock decision.
RESULTS: The sensitivity and PPV of the pause detector were 93.5% and 97.3%, respectively. The sensitivity and specificity of the SAA in the detected pauses were 93.8% (90% low CI, 90.0%) and 95.9% (90% low CI, 94.7%), respectively. Using the method, shocks would have been advanced in 97% of occasions. For patients in nonshockable rhythms, rhythm reassessment pauses would be avoided in 95.2% (95% CI, 91.6-98.8) of occasions, thus increasing the overall chest compression fraction (CCF).
CONCLUSION: An automatic method could be used to safely analyse the rhythm during ventilation pauses. This would contribute to an early detection of refibrillation, and to increase CCF in patients with nonshockable rhythms.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app