Applied Radiation and Isotopes 2015 March
INTRODUCTION: High-yielding, automated production of a PET tracer that reflects proliferation, 3'-deoxy-3'-[(18)F]fluorothymidine ([(18)F]FLT), is reported using a modified Bioscan Coincidence FDG reaction module.
METHODS: Production of [(18)F]FLT was implemented through: (1) modification of an original FDG manifold; (2) application of an alternate time sequence; and (3) altered solid-phase extraction (SPE) purification. Quality control testing, including standard radiochemical figures of merit and preclinical positron emission tomography (PET) imaging, was carried out.
RESULTS: High decay-corrected yields of [(18)F]FLT (16-39%) were reproducibly obtained. The product exhibited very high specific activity (4586.9TBq/mmol; 123,969Ci/mmol) and radiochemical purity (>99%). Overall, the [(18)F]FLT produced in this manner was superior to typical productions that utilized a GE TRACERlab FXF-N reaction module. Additionally, purification with SPE cartridges, followed by manual elution, accelerated overall run time and resulted in a two-fold increase in [(18)F]FLT concentration. PET imaging showed the [(18)F]FLT produced by this method was highly suitable for non-invasive tumor imaging in mice.
CONCLUSIONS: The Bioscan Coincidence GE FDG Reaction Module was readily adapted to reproducibly provide [(18)F]FLT in high yield, specific activity, and radiochemical purity. The approach was suitable to provide sufficient amounts of material for preclinical studies.
METHODS: Production of [(18)F]FLT was implemented through: (1) modification of an original FDG manifold; (2) application of an alternate time sequence; and (3) altered solid-phase extraction (SPE) purification. Quality control testing, including standard radiochemical figures of merit and preclinical positron emission tomography (PET) imaging, was carried out.
RESULTS: High decay-corrected yields of [(18)F]FLT (16-39%) were reproducibly obtained. The product exhibited very high specific activity (4586.9TBq/mmol; 123,969Ci/mmol) and radiochemical purity (>99%). Overall, the [(18)F]FLT produced in this manner was superior to typical productions that utilized a GE TRACERlab FXF-N reaction module. Additionally, purification with SPE cartridges, followed by manual elution, accelerated overall run time and resulted in a two-fold increase in [(18)F]FLT concentration. PET imaging showed the [(18)F]FLT produced by this method was highly suitable for non-invasive tumor imaging in mice.
CONCLUSIONS: The Bioscan Coincidence GE FDG Reaction Module was readily adapted to reproducibly provide [(18)F]FLT in high yield, specific activity, and radiochemical purity. The approach was suitable to provide sufficient amounts of material for preclinical studies.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app