Comparative Study
Journal Article
Research Support, N.I.H., Extramural
Add like
Add dislike
Add to saved papers

Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging.

PURPOSE: The objective of this study was to evaluate the performance of the built-in MR-based attenuation correction (MRAC) included in the combined whole-body Ingenuity TF PET/MR scanner and compare it to the performance of CT-based attenuation correction (CTAC) as the gold standard.

METHODS: Included in the study were 26 patients who underwent clinical whole-body FDG PET/CT imaging and subsequently PET/MR imaging (mean delay 100 min). Patients were separated into two groups: the alpha group (14 patients) without MR coils during PET/MR imaging and the beta group (12 patients) with MR coils present (neurovascular, spine, cardiac and torso coils). All images were coregistered to the same space (PET/MR). The two PET images from PET/MR reconstructed using MRAC and CTAC were compared by voxel-based and region-based methods (with ten regions of interest, ROIs). Lesions were also compared by an experienced clinician.

RESULTS: Body mass index and lung density showed significant differences between the alpha and beta groups. Right and left lung densities were also significantly different within each group. The percentage differences in uptake values using MRAC in relation to those using CTAC were greater in the beta group than in the alpha group (alpha group -0.2 ± 33.6%, R(2) = 0.98, p < 0.001; beta group 10.31 ± 69.86%, R(2) = 0.97, p < 0.001).

CONCLUSION: In comparison to CTAC, MRAC led to underestimation of the PET values by less than 10% on average, although some ROIs and lesions did differ by more (including the spine, lung and heart). The beta group (imaged with coils present) showed increased overall PET quantification as well as increased variability compared to the alpha group (imaged without coils). PET data reconstructed with MRAC and CTAC showed some differences, mostly in relation to air pockets, metallic implants and attenuation differences in large bone areas (such as the pelvis and spine) due to the segmentation limitation of the MRAC method.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app