CT angiography of the neck: value of contrast medium dose reduction with low tube voltage and high tube current in a 64-detector row CT.
Clinical Radiology 2014 April
AIM: To evaluate the feasibility of a low-dose contrast medium protocol for 64-detector row computed tomography angiography (CTA) of the neck using a low-tube-voltage/high-tube-current setting.
MATERIALS AND METHODS: A phantom study was performed using 64-detector row spiral CT at multiple tube voltage and current settings. Iodine contrast medium attenuation curves were acquired by processing and used to select the best contrast medium-to-noise ratio (CNR). A prospective clinical study was then performed on 84 patients requiring neck CTA. Patients were randomly divided into two groups of 42. Group A was examined using the conventional imaging protocol (120 kV, 400 mAs) and group B was examined at 80 kV and 600 mAs along with a 50% reduction in contrast medium dose. The CT dose index-volume (CTDI(vol)), background noise (BN), and CNR were measured and statistically analysed. Various image quality criteria were evaluated by two senior radiologists using a qualitative five-point scale.
RESULTS: Comparing group B with A, CTDIvol decreased by 54% (B: 27.48 mGy, A: 59.11 mGy), however, the CNR increased by 50%. The mean attenuation, which was caused by venous streak artefacts, was significantly lower in group B than A. Qualitative image analysis found that all criteria were significantly better for group B than A.
CONCLUSION: At 64-detector row spiral CT, the low-tube-voltage/high-tube-current with low-dose contrast medium protocol was superior to the conventional protocol regarding radiation dose, venous streak artefacts, and image quality, and is feasible for CTA of the neck.
MATERIALS AND METHODS: A phantom study was performed using 64-detector row spiral CT at multiple tube voltage and current settings. Iodine contrast medium attenuation curves were acquired by processing and used to select the best contrast medium-to-noise ratio (CNR). A prospective clinical study was then performed on 84 patients requiring neck CTA. Patients were randomly divided into two groups of 42. Group A was examined using the conventional imaging protocol (120 kV, 400 mAs) and group B was examined at 80 kV and 600 mAs along with a 50% reduction in contrast medium dose. The CT dose index-volume (CTDI(vol)), background noise (BN), and CNR were measured and statistically analysed. Various image quality criteria were evaluated by two senior radiologists using a qualitative five-point scale.
RESULTS: Comparing group B with A, CTDIvol decreased by 54% (B: 27.48 mGy, A: 59.11 mGy), however, the CNR increased by 50%. The mean attenuation, which was caused by venous streak artefacts, was significantly lower in group B than A. Qualitative image analysis found that all criteria were significantly better for group B than A.
CONCLUSION: At 64-detector row spiral CT, the low-tube-voltage/high-tube-current with low-dose contrast medium protocol was superior to the conventional protocol regarding radiation dose, venous streak artefacts, and image quality, and is feasible for CTA of the neck.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app