Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality.

PURPOSE: To evaluate the effect of a cardiac bowtie-shaped filter in an ultra high pitch CTPA protocol at 100 kV on image quality and radiation dose.

MATERIALS AND METHODS: Retrospective study of 100 patients referred for CTPA. 50 patients scanned with a standard 100 kV protocol at pitch 2.8 (Protocol A) and 50 patients scanned with a 100 kV protocol at pitch 3.2 with a cardiac bowtie-shaped filter (Protocol B). All other scanning parameters kept constant. Images from both groups reconstructed with filtered back projection and iterative reconstruction. Central pulmonary vessel attenuation and background noise were quantitatively measured and signal-to-noise (SNR) and contrast-to-noise (CNR) were calculated. Two radiologists performed qualitative assessment grading visualization of the pulmonary vasculature and noise level. CTDIvol and DLP were recorded and effective dose was calculated.

RESULTS: CTDIvol, DLP and effective dose were significantly (p<0.0001) lower in Protocol B (2.3 ± 0.5 mGy, 78.4 ± 16.5 mGycm, 1.4 ± 0.3 mSy, respectively) compared to Protocol A (4.3 ± 0.5 mGy, 152.0 ± 19.6 mGycm, 2.7 ± 0.3 mSy, respectively). Protocol B had significantly (p<0.0001) higher noise than Protocol A (23.8 ± 6.9 HU vs 36.8 ± 7.3 HU) and lower SNR (11.8 ± 3.7 HU vs 19.2 ± 8.1 HU) and CNR (10.3 ± 3.7 HU vs 24.9 ± 13.4 HU) but there was no significant difference in the subjective visualization of the pulmonary vasculature (p=0.63). Furthermore, iterative reconstruction significantly (p<0.0001) improves image noise (29.4 ± 5.5 HU from 36.8 ± 7.3 HU).

CONCLUSION: The addition of a cardiac bowtie-shaped filter with an ultra high pitch CTPA protocol at 100 kV resulted in a 48% dose reduction without significantly affecting diagnostic image quality. In addition, the use of iterative reconstruction significantly improves image quality by reducing noise permitting the possibility for further dose reduction strategies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app