Predicting the postoperative intraocular lens position using continuous intraoperative optical coherence tomography measurements.
Investigative Ophthalmology & Visual Science 2013 August
PURPOSE: The aim of this study was to assess, if measuring the position of the lens capsule intraoperatively with a continuous intraoperative optical coherence tomography (OCT) device could be useful to improve the prediction of the intraocular lens (IOL) position.
METHODS: This prospective study included patients who were scheduled for cataract surgery. A prototype operating microscope with an integrated continuous OCT device was used to measure the anterior and posterior lens capsule position at different time points during cataract surgery. In all cases, a capsular tension ring (CTR) was used to tauten the lens capsule. Partial coherence interferometry was used to measure anterior chamber depth (ACD) immediately before, and 1 hour and 3 months postoperatively. Partial least squares regression (PLSR) was used to assess the influence of different pre- and intraoperatively measured parameters.
RESULTS: In total, 70 eyes of 70 patients were included. Mean axial eye length was 23.6 mm (range, 20.6 mm-30.8 mm), mean used IOL power was 22.2 diopters (D; range, 6.0 D-31.5 D). PLSR showed that the anterior lens capsule measured after removing the crystalline lens and after implanting a CTR was a significantly better predictor for the postoperative ACD compared with preoperative ACD measurements.
CONCLUSIONS: The main problem of IOL power calculation, the prediction of the IOL position after surgery, could possibly be reduced by using intraoperative lens capsule measurements instead of preoperative ACD measurements. (ClinicalTrials.gov number, NCT01867541.).
METHODS: This prospective study included patients who were scheduled for cataract surgery. A prototype operating microscope with an integrated continuous OCT device was used to measure the anterior and posterior lens capsule position at different time points during cataract surgery. In all cases, a capsular tension ring (CTR) was used to tauten the lens capsule. Partial coherence interferometry was used to measure anterior chamber depth (ACD) immediately before, and 1 hour and 3 months postoperatively. Partial least squares regression (PLSR) was used to assess the influence of different pre- and intraoperatively measured parameters.
RESULTS: In total, 70 eyes of 70 patients were included. Mean axial eye length was 23.6 mm (range, 20.6 mm-30.8 mm), mean used IOL power was 22.2 diopters (D; range, 6.0 D-31.5 D). PLSR showed that the anterior lens capsule measured after removing the crystalline lens and after implanting a CTR was a significantly better predictor for the postoperative ACD compared with preoperative ACD measurements.
CONCLUSIONS: The main problem of IOL power calculation, the prediction of the IOL position after surgery, could possibly be reduced by using intraoperative lens capsule measurements instead of preoperative ACD measurements. (ClinicalTrials.gov number, NCT01867541.).
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app