Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand.

PURPOSE: This study was carried out to assess the radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Bangkok, Thailand.

METHODS: Data were collected from 60 patients under transarterial oily-chemoembolisation (TOCE) and femoral angiography performed with the Toshiba Infinix model VC-i FPD single plane system. Data were also collected from 60 patients who underwent brain arteriovenous malformations (AVM) and dural-arteriovenous fistula (DAVF) embolisation, performed with the Toshiba Infinix model VF-i bi-plane systems. A built-in air kerma area product (KAP) meter calibrated in situ was used for the skin dose calculation.

RESULTS: The calibration coefficient of air kerma area product meter at tube voltage between 50 kV and 100 kV was found to vary within ± 5.07%, ± 7.2%, ± 4.86 % from calibration coefficient of 80 kV for a single-plane, tube 1 and tube 2 of bi-plane x-ray system, respectively. Mean air kerma area product values were 90.99 ± 52.89, 31.02 ± 17.92, 33.11 ± 23.99 (Frontal), 35.01 ± 19.10 (Lateral), 50.15 ± 44.76 (Frontal), 97.31 ± 44.12 (Lateral) Gy-cm(2) for transarterial oily-chemoembolisation, femoral angiography, diagnostic cerebral angiography, therapeutic cerebral angiography, respectively. The therapeutic cerebral angiography procedure was found to give the highest entrance dose, number of images and fluoroscopy time: 362.63 cGy (Lateral), 1015 images (Lateral) and 126 minutes, respectively. However, the highest air kerma area product value was from transarterial oily-chemoembolisation with 264.37 Gy-cm(2). There were 2 cases of therapeutic cerebral angiography, where the patient entrance dose was higher than 3 Gy in the frontal view, which reached the deterministic threshold for temporary epilation.

CONCLUSION: Very wide variationswere found in patient dose from different interventional procedures. There is a need for a dose record system to provide feedback to radiologists who perform the procedures; especially in cases where the dose exceeds the deterministic threshold.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app