Cardiac output determination from endotracheally measured impedance cardiography: clinical evaluation of endotracheal cardiac output monitor.

OBJECTIVES: To evaluate the accuracy, precision, and trending of a new endotracheally sourced impedance cardiography-based cardiac output (CO) monitor (ECOM; ConMed Corp, Irvine, CA).

SETTING: Two university hospitals.

PARTICIPANTS: Thirty patients scheduled for elective coronary artery bypass graft (CABG) surgery.

INTERVENTIONS: All patients received a pulmonary artery catheter (PAC), arterial catheter, endotracheal CO monitor (ECOM), endotracheal intubation, and transesophageal echocardiographic monitoring. ECOM CO was compared with CO measured with pulmonary artery thermodilution, and left ventricular CO measured with transesophageal echocardiography.

MEASUREMENTS: One hundred forty-five pairs of triplicate CO measurements using intermittent bolus pulmonary artery thermodilution (TD) and ECOM were compared at 5 distinct time points: postinduction, postinduction passive leg raise, poststernotomy, post-CABG completion, and post-chest closure. Eighty-seven pairs of triplicate CO measurements using transesophageal echocardiography were obtained at 3 time points: postinduction, post-CABG completion, and post-chest closure and compared with ECOM- and PA-derived CO measurements. The measurements at each time point were compared by using Bland-Altman and polar plot analyses.

RESULTS: The mean CO ranged from 2.16 to 9.41 L/min. ECOM CO, compared with TD CO, revealed a bias of 0.02 L/min, 95% limits of agreement of -2.26 to 2.30 L/min, and a percent error of 50%. ECOM CO showed trending with TD CO with 91% and 99% of values within 0.5L/min and 1 L/min limits of agreement, respectively. ECOM CO, compared with TEE CO, revealed a bias of -0.25 L/min, 95% limits of agreement of -2.41 to 1.92 L/min, and a percent error of 48%. ECOM CO showed trending with TEE CO with 83% and 95% of values within 0.5L/min and 1 L/min limits of agreement, respectively.

CONCLUSION: ECOM CO shows an acceptable bias with wide limits of agreement and a large percent error when compared with TD CO or TEE CO; however, it shows acceptable trending of CO to both modalities in patients undergoing cardiac surgery. Further studies are required to evaluate ECOM in other patient populations and clinical situations.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app