BACKGROUND AND METHODS: the most reliable methods for confirming endotracheal tube placement are direct visualization of passage through the vocal cords and documentation of CO2 in the expired gas. We evaluated the use of a disposable colorimetric CO2 detector for verifying endotracheal tube position in small animals. The end-tidal CO2 (Petco2) detector was tested in 11 piglets with the endotracheal tube sequentially in the trachea, the esophagus, the esophagus with a carbonated beverage in the stomach, the esophagus after bag-mask ventilation. Endotracheal tube position was confirmed in all cases by direct visualization and capnometry.

RESULTS: The Petco2 detector identified the tube placement accurately in all 54 (21 tracheal, 33 esophageal) intubations (p less than .001).

CONCLUSIONS: This disposable Petco2 detector is highly sensitive and specific for verifying endotracheal tube placement in this nonarrest piglet model.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app