We have located links that may give you full text access.
Is visual assessment of jaundice reliable as a screening tool to detect significant neonatal hyperbilirubinemia?
Journal of Pediatrics 2008 June
OBJECTIVE: To assess the reliability of visual assessment of bilirubin levels (BiliEye) in newborns as a screening tool to detect significant neonatal hyperbilirubinemia.
STUDY DESIGN: 5 neonatologists and 17 nurses estimated 3,532 BiliEye in 1,129 term and late preterm (> or = 35 weeks) infants before discharge from the nursery, at 62 +/- 24 hours. Total serum bilirubin (TSB) levels were measured concomitantly.
RESULTS: Mean TSB and BiliEye were 6.7 +/- 2.9 mg/dL (range, 0.4-18.2 mg/dL) and 6.6 +/- 3.2 mg/dL (range, 0.0-17.2 mg/dL), respectively, with good correlation (Pearson's r = 0.752, P < .0001), but other measures of agreement were poor. 61.5% of the 109 babies with TSB levels in high-risk zones were clinically misclassified. The area under curve (AUC) of the receiver-operating characteristics plotted for these high-risk zones was 0.825, but became low for early discharge (< or = 36 hours; AUC = 0.638) and late preterm (35-37 weeks; AUC = 0.613). There was significant interobserver variation (low weighted kappa, 0.363).
CONCLUSIONS: Although there was good correlation between BiliEye and actual TSB level, visual assessment was unreliable as a screening tool to detect significant neonatal hyperbilirubinemia before discharge. Babies with TSB levels within high-risk zones may be clinically misdiagnosed as low-risk, resulting in inadequate follow-up.
STUDY DESIGN: 5 neonatologists and 17 nurses estimated 3,532 BiliEye in 1,129 term and late preterm (> or = 35 weeks) infants before discharge from the nursery, at 62 +/- 24 hours. Total serum bilirubin (TSB) levels were measured concomitantly.
RESULTS: Mean TSB and BiliEye were 6.7 +/- 2.9 mg/dL (range, 0.4-18.2 mg/dL) and 6.6 +/- 3.2 mg/dL (range, 0.0-17.2 mg/dL), respectively, with good correlation (Pearson's r = 0.752, P < .0001), but other measures of agreement were poor. 61.5% of the 109 babies with TSB levels in high-risk zones were clinically misclassified. The area under curve (AUC) of the receiver-operating characteristics plotted for these high-risk zones was 0.825, but became low for early discharge (< or = 36 hours; AUC = 0.638) and late preterm (35-37 weeks; AUC = 0.613). There was significant interobserver variation (low weighted kappa, 0.363).
CONCLUSIONS: Although there was good correlation between BiliEye and actual TSB level, visual assessment was unreliable as a screening tool to detect significant neonatal hyperbilirubinemia before discharge. Babies with TSB levels within high-risk zones may be clinically misdiagnosed as low-risk, resulting in inadequate follow-up.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app