The reduction of endplate fractures during balloon vertebroplasty: a detailed radiological analysis of the treatment of burst fractures using pedicle screws, balloon vertebroplasty, and calcium phosphate cement.

Spine 2005 August 16
STUDY DESIGN: In a human cadaveric burst fracture model, the amount of endplate fracture reduction after posterior instrumentation and balloon vertebroplasty was investigated quantitatively.

OBJECTIVES: To assess, in a burst fracture model, the vertebral body and adjacent disc heights, in parallel sagittal planes with 3-dimensional (3D) rotational x-ray imaging, at various phases during pedicle screw fixation and subsequent balloon vertebroplasty.

SUMMARY OF BACKGROUND DATA: In recent human cadaveric thoracolumbar fracture studies, it was found that vertebral body height could be restored significantly with inflatable bone tamps. However, limited quantitative data exist on the amount of fracture reduction that can be achieved and how much of the reduction will be lost after deflation and removal of the bone tamps before the cement is injected.

METHODS: Twenty burst fractures were created and balloon vertebroplasty with calcium phosphate cement was performed after pedicle screw instrumentation. A 3D dataset was obtained during the following phases: intact, fractured, after reduction and stabilization with pedicle screws, after inflation of the balloons, after deflation and removal of the balloons, after injection of the cement. The fractured vertebral body and adjacent disc heights were measured from five reconstructed sagittal images and compared for the six phases of the procedure. Furthermore, the difference between the vertebral body height centrally and peripherally was calculated.

RESULTS: The mean vertebral body height at the thoracic level was Tintact = 19.5 +/- 2.2 mm, Tfractured = 14.6 +/- 3.8 mm, Treduction = 17.3 +/- 2.2 mm, Tinflation = 20.1 +/- 2.0 mm, Tdeflation = 18.0 +/- 2.0 mm, and Tcement = 17.8 +/- 1.8 mm. The overall change in vertebral body height between these phases was significant (P < 0.001). At the lumbar level the mean vertebral body height was Tintact = 23.2 +/- 3.8 mm, Tfractured = 14.7 +/- 3.0 mm, Treduction = 18.4 +/- 2.5 mm, Tinflation = 23.2 +/- 3.5 mm, Tdeflation = 19.3 +/- 2.3 mm, and Tcement = 20.2 +/- 2.8 mm. The overall change in MCVBH between these phases was also significant (P < 0.001). The increase in vertebral body height resulted in a decrease of the adjacent disc height. No difference was found for the amount of endplate reduction in the center or at the periphery. No leakage of cement was detected in the spinal canal.

CONCLUSIONS: Reduction of endplate fractures, both in the center and at the periphery, seems feasible and safe with combined fracture reduction and balloon vertebroplasty. The endplate fracture reduction that was gained by inflation of the bone tamps could not be maintained after deflation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app