JOURNAL ARTICLE

Effect of different inspiratory rise time and cycling off criteria during pressure support ventilation in patients recovering from acute lung injury

Davide Chiumello, Paolo Pelosi, Paolo Taccone, Arthur Slutsky, Luciano Gattinoni
Critical Care Medicine 2003, 31 (11): 2604-10
14605531

OBJECTIVE: With many mechanical ventilators, it is possible to modify the time to reach the selected airway pressure and the criteria for cycling off the inflation during pressure support ventilation. This study evaluated the effect of different inspiratory rise time and cycling off criteria on breathing pattern and work of breathing.

DESIGN: Clinical study.

SETTING: University laboratory.

PATIENTS: Ten intubated patients recovering from acute lung injury (PaO2/FiO2 245 +/- 26 torr, positive end-expiratory pressure 9 +/- 3 cm H2O).

INTERVENTIONS: We studied two inspiratory rise time criteria (shortest and longest, 0% and 40% of the breath cycle time) and two cycling off criteria (lowest and highest, 5% and 40% of the peak inspiratory flow) at 5 and 15 cm H2O of pressure support. Respiratory rate, tidal volume, and inspiratory and expiratory work of breathing (WOBI and WOBE) were measured.

MEASUREMENTS AND MAIN RESULTS: At both levels of pressure support ventilation, the shortest inspiratory rise time significantly reduced the WOBI from 0.77 +/- 0.32 to 0.56 +/- 0.23 J/L and from 0.24 +/- 0.28 to 0.08 +/- 0.09 J/L without affecting respiratory rate or tidal volume. At 15 cm H2O of pressure support ventilation, the lowest cycling off criteria significantly reduced respiratory rate from 24.9 +/- 12.1 to 21.5 +/- 12.7 beats/min and increased tidal volume from 0.51 +/- 0.17 to 0.60 +/- 0.26 L. At both levels of pressure support ventilation, the modification of cycling off criteria did not influence WOBI and WOBE.

CONCLUSIONS: Our results suggest that in patients recovering from acute lung injury during pressure support ventilation, a) the shortest inspiratory rise time reduces the WOBI; and b) at 15 cm H2O of pressure support ventilation, the lowest cycling off criteria reduces the respiratory rate and increases the tidal volume without modifying the WOBI and WOBE. Modifications of inspiratory rise time and cycling off criteria must be carefully adjusted during pressure support ventilation.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
14605531
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"