Effect of different inspiratory rise time and cycling off criteria during pressure support ventilation in patients recovering from acute lung injury.

OBJECTIVE: With many mechanical ventilators, it is possible to modify the time to reach the selected airway pressure and the criteria for cycling off the inflation during pressure support ventilation. This study evaluated the effect of different inspiratory rise time and cycling off criteria on breathing pattern and work of breathing.

DESIGN: Clinical study.

SETTING: University laboratory.

PATIENTS: Ten intubated patients recovering from acute lung injury (PaO2/FiO2 245 +/- 26 torr, positive end-expiratory pressure 9 +/- 3 cm H2O).

INTERVENTIONS: We studied two inspiratory rise time criteria (shortest and longest, 0% and 40% of the breath cycle time) and two cycling off criteria (lowest and highest, 5% and 40% of the peak inspiratory flow) at 5 and 15 cm H2O of pressure support. Respiratory rate, tidal volume, and inspiratory and expiratory work of breathing (WOBI and WOBE) were measured.

MEASUREMENTS AND MAIN RESULTS: At both levels of pressure support ventilation, the shortest inspiratory rise time significantly reduced the WOBI from 0.77 +/- 0.32 to 0.56 +/- 0.23 J/L and from 0.24 +/- 0.28 to 0.08 +/- 0.09 J/L without affecting respiratory rate or tidal volume. At 15 cm H2O of pressure support ventilation, the lowest cycling off criteria significantly reduced respiratory rate from 24.9 +/- 12.1 to 21.5 +/- 12.7 beats/min and increased tidal volume from 0.51 +/- 0.17 to 0.60 +/- 0.26 L. At both levels of pressure support ventilation, the modification of cycling off criteria did not influence WOBI and WOBE.

CONCLUSIONS: Our results suggest that in patients recovering from acute lung injury during pressure support ventilation, a) the shortest inspiratory rise time reduces the WOBI; and b) at 15 cm H2O of pressure support ventilation, the lowest cycling off criteria reduces the respiratory rate and increases the tidal volume without modifying the WOBI and WOBE. Modifications of inspiratory rise time and cycling off criteria must be carefully adjusted during pressure support ventilation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app