Fluoroscopically calibrated 3D-printed patient-specific instruments improve the accuracy of osteotomy during bone tumor resection adjacent to joints.

BACKGROUND: Inadequate surface matching, variation in the guide design, and soft tissue on the skeletal surface may make it difficult to accurately place the 3D-printed patient-specific instrument (PSI) exactly to the designated site, leading to decreased accuracy, or even errors. Consequently, we developed a novel 3D-printed PSI with fluoroscopy-guided positioning markers to enhance the accuracy of osteotomies in joint-preserving surgery. The current study was to compare whether the fluoroscopically calibrated PSI (FCPSI) can achieve better accuracy compared with freehand resection and conventional PSI (CPSI) resection.

METHODS: Simulated joint-preserving surgery was conducted using nine synthetic left knee bone models. Osteotomies adjacent to the knee joint were designed to evaluate the accuracy at the epiphysis side. The experiment was divided into three groups: free-hand, conventional PSI (CPSI), and fluoroscopically Calibrated PSI (FCPSI). Post-resection CT scans were quantitatively analyzed. Analysis of variance (ANOVA) was used.

RESULT: FCPSI improved the resection accuracy significantly. The mean location accuracy is 2.66 mm for FCPSI compared to 6.36 mm (P < 0.001) for freehand resection and 4.58 mm (P = 0.012) for CPSI. The mean average distance is 1.27 mm compared to 2.99 mm (p < 0.001) and 2.11 mm (p = 0.049). The mean absolute angle is 2.16° compared to 8.50° (p < 0.001) and 5.54° (p = 0.021). The mean depth angle is 1.41° compared to 8.10° (p < 0.001) and 5.32° (p = 0.012). However, there were no significant differences in the front angle compared to the freehand resection group (P = 0.055) and CPSI (P = 0.599) group. The location accuracy observed with FCPSI was maintained at 4 mm, while CPSI and freehand resection exhibited a maximum deviation of 8 mm.

CONCLUSION: The fluoroscopically calibrated 3D-printed patient-specific instruments improve the accuracy of osteotomy during bone tumor resection adjacent to joint joints compared to conventional PSI and freehand resection. In conclusion, this novel 3D-printed PSI offers significant accuracy improvement in joint preserving surgery with a minimal increase in time and design costs.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app