Single suture-mediated closure system after transfemoral transcatheter aortic valve implantation: A single-center real-world experience.
Catheterization and Cardiovascular Interventions 2024 April 20
BACKGROUND: Despite the use of two crossed Perclose ProGlide™ (Abbott Vascular Devices) is the most widespread technique to close the main arterial access in transfemoral transcatheter aortic valve implantation (TF-TAVI), the safest and most effective strategy still remains much debated.
AIMS: The aim of the present study was to evaluate the performance of a single Perclose ProGlide suture-mediated closure device to obtain femoral hemostasis after sheathless implantation of self-expanding transcatheter heart valves through their 14 F-equivalent fix delivery systems.
METHODS: This prospective observational study included 439 patients undergoing TF-TAVI at the "Montevergine" Clinic of Mercogliano, Italy. All patients underwent hemostasis of the large-bore access using a single Perclose ProGlide with preclose technique, after sheathless implantation of self-expanding transcatheter heart valves through 14 F-equivalent fix delivery systems. A multidetector computed tomography analysis of size, tortuosity, atherosclerotic, and calcification burdens of the ilio-femoral access route was made by a dedicated corelab. Vascular complications (VCs), percutaneous closure device (PCD) failure, and bleedings were adjudicated by a clinical events committee.
RESULTS: A total of 81 different VCs were observed in 60 patients (13.7%); among these, 41 (5% of patients) were categorized as major. PCD failure occurred in 14 patients (3.2%). At the logistic regression analysis, no predictors of PCD failure have been identified.
CONCLUSION: This registry suggests that the use of a single suture-mediated closure device could be considered a safe and efficient technique to achieve access site hemostasis in patients undergoing TF-TAVI through 14 F-equivalent fix delivery systems.
AIMS: The aim of the present study was to evaluate the performance of a single Perclose ProGlide suture-mediated closure device to obtain femoral hemostasis after sheathless implantation of self-expanding transcatheter heart valves through their 14 F-equivalent fix delivery systems.
METHODS: This prospective observational study included 439 patients undergoing TF-TAVI at the "Montevergine" Clinic of Mercogliano, Italy. All patients underwent hemostasis of the large-bore access using a single Perclose ProGlide with preclose technique, after sheathless implantation of self-expanding transcatheter heart valves through 14 F-equivalent fix delivery systems. A multidetector computed tomography analysis of size, tortuosity, atherosclerotic, and calcification burdens of the ilio-femoral access route was made by a dedicated corelab. Vascular complications (VCs), percutaneous closure device (PCD) failure, and bleedings were adjudicated by a clinical events committee.
RESULTS: A total of 81 different VCs were observed in 60 patients (13.7%); among these, 41 (5% of patients) were categorized as major. PCD failure occurred in 14 patients (3.2%). At the logistic regression analysis, no predictors of PCD failure have been identified.
CONCLUSION: This registry suggests that the use of a single suture-mediated closure device could be considered a safe and efficient technique to achieve access site hemostasis in patients undergoing TF-TAVI through 14 F-equivalent fix delivery systems.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app