Interchanging Reusable and Disposable Nebulizers Used with Home-Based Compressors May Result in Inconsistent Dosing: A Laboratory Investigation with Device Combinations Supplied to the US Healthcare Environment.

Pulmonary Therapy 2024 April 16
INTRODUCTION: Reusable nebulizer-compressor combinations deliver inhaled medications for patients with chronic lung diseases. On hospital discharge, the patient may take home the disposable nebulizer that was packaged and combine it with their home compressor. Though this practice may reduce waste, it can increase variability in medication delivery. Our study compared several reusable and disposable nebulizers packaged with compressor kits used in the US. We included a common disposable hospital nebulizer that may not be supplied with popular home kits but may be brought home after a hospitalization or emergency department visit. We focused on fine droplet mass < 4.7 μm aerodynamic diameter (FDM<4.7 μm ), associated with medication delivery to the airways of the lungs.

METHODS: We evaluated the following nebulizer-compressor combinations (n = 5 replicates): 1. OMBRA® Table Top Compressor with MC 300® reusable and Airlife™ MistyMax™ 10® disposable nebulizer, 2. Sami-the-Seal® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 3. VIOS® compressor with LC Sprint® reusable, and VixOne® and Airlife™ MistyMax™ disposable nebulizers, 4. Innospire® Elegance® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 5. Willis-the-Whale® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 6. Pari PRONEB® Max compressor with LC Sprint® reusable and Airlife™ MistyMax 10™ disposable nebulizer. We placed a 3-ml albuterol solution (0.833 mg/ml) in each nebulizer. A bacterial/viral filter was attached to the nebulizer mouthpiece to capture emitted medication, with the filter exit coupled to a simulator of a tidal breathing adult (rate = 10 cycles/min; Vt  = 600 ml; I/E ratio = 1:2). The filter was replaced at 1-min intervals until onset of sputter. Droplet size distributions (n = 5 replicates/system) were determined in parallel by laser diffractometry.

RESULTS: Cumulative FDM<4.7 μm varied from 381 ± 33 μg for the best performing combination (Proneb/LC-Sprint) to 150 ± 21 μg for the system with the lowest output (VIOS®/MistyMax 10™).

CONCLUSIONS: Substituting one nebulizer for another can result in large differences in medication delivery to the lungs.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app