Measurements regarding a combined therapy concept for ophthalmic tumors consisting of brachytherapy and x-rays.

Objective
We present a novel concept to treat ophthalmic tumors which combines brachytherapy and low-energy x-ray therapy. Brachytherapy with106 Ru applicators applicators is inadequate for intraocular tumors with a height of 7 mm and more. This results from a steep dose gradient, and it is unfeasible to deliver the required dose at the tumor apex without exceeding the maximum tolerable sclera dose of usually 1000 to 1500 Gy. Other modalities, such as irradiation with charged particles, may be individually contraindicated. A dose boost at the apex provided by an x-ray therapy unit, performed simultaneously with the brachytherapy, results in a more homogeneous dose profile than brachytherapy alone. This avoids damage to organs at risk. The applicator may also serve as a beam stop for x-rays passing through the target volume, which reduces healthy tissue dosage. This study aims to investigate the suitability of the applicator to serve as a beam stop for the x-rays. 

Approach
A phantom with three detector types comprising a soft x-ray ionization chamber, radiochromic films, and a self-made scintillation detector was constructed to perform dosimetry. Measurements were performed using a conventional x-ray unit for superficial therapy to investigate the uncertainties of the phantom and the ability of the applicator to absorb x-rays. The manufacturer provided a dummy plaque to obtain x-ray dose profiles without noise from106 Ru decays.

Results
The phantom is generally feasible to obtain dose profiles with three different detector types. The interaction of x-rays with the silver of the applicator leads to an increased dose rate in front of the applicator. The dose rate of the x-rays is reduced by up to 90 % behind a106 Ru applicator. Therefore, a106 Ru applicator can be used as a beam stop in combined x-ray and brachytherapy treatment.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app