We have located links that may give you full text access.
Accuracy and Reproducibility of Virtual Reality Perimetry in Children.
Journal of Pediatric Ophthalmology and Strabismus 2024 Februrary 19
PURPOSE: To determine normal threshold values for the 24-2 program of the VisuALL-K (Olleyes, Inc), a portable static automated virtual reality perimeter that uses a video game-based testing modality on an updated version of the algorithm and to repeat the test to assess retest variability.
METHODS: This prospective cohort study included 19 participants (aged 8 to 17 years) with healthy afferent systems. Participants performed two 24-2 automated perimetry tests on the virtual reality perimeter device.
RESULTS: Mean age was 12.7 ± 2.5 years; 53% were boys and 47% were girls. Mean test location threshold was 30.5 ± 4.0 dB for test 1 and 31.3 ± 2.8 dB for test 2. Sensitivity decreased with eccentricity as expected (-0.17 dB per degree). Bland-Altman analysis showed a repeatability coefficient of ±5.1 dB for mean deviation, meaning that 95% of retest values were within ±5 dB of the threshold value for the first test. Higher thresholds for the first test regressed toward the mean values in the second test, indicating a learning effect. Intraclass correlation coefficient for mean deviation between test 1 and test 2 was 0.70 (0.55 to 0.80, P < .0001). Mean time to completion for the virtual reality perimeter was 5.1 ± 1.5 min/eye.
CONCLUSIONS: The VisuALL-K virtual reality perimeter provided an assessment of the visual field in normal children aged 8 to 17 years. A shallow hill of vision was demonstrated. A notable learning effect was demonstrated, with repeatability difference wider than other published head-mounted display perimeters. [ J Pediatr Ophthalmol Strabismus . 20XX:X(X):XX-XX.] .
METHODS: This prospective cohort study included 19 participants (aged 8 to 17 years) with healthy afferent systems. Participants performed two 24-2 automated perimetry tests on the virtual reality perimeter device.
RESULTS: Mean age was 12.7 ± 2.5 years; 53% were boys and 47% were girls. Mean test location threshold was 30.5 ± 4.0 dB for test 1 and 31.3 ± 2.8 dB for test 2. Sensitivity decreased with eccentricity as expected (-0.17 dB per degree). Bland-Altman analysis showed a repeatability coefficient of ±5.1 dB for mean deviation, meaning that 95% of retest values were within ±5 dB of the threshold value for the first test. Higher thresholds for the first test regressed toward the mean values in the second test, indicating a learning effect. Intraclass correlation coefficient for mean deviation between test 1 and test 2 was 0.70 (0.55 to 0.80, P < .0001). Mean time to completion for the virtual reality perimeter was 5.1 ± 1.5 min/eye.
CONCLUSIONS: The VisuALL-K virtual reality perimeter provided an assessment of the visual field in normal children aged 8 to 17 years. A shallow hill of vision was demonstrated. A notable learning effect was demonstrated, with repeatability difference wider than other published head-mounted display perimeters. [ J Pediatr Ophthalmol Strabismus . 20XX:X(X):XX-XX.] .
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2025 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app