Refined scan protocol for the evaluation of pulmonary perfusion standardized image quality and reduced radiation dose in dynamic chest radiography.

PURPOSE: Dynamic chest radiography (DCR) is a novel imaging technique used to noninvasively evaluate pulmonary perfusion. However, the standard DCR protocol, which is roughly adapted to the patient's body size, occasionally causes over- or underexposure, which could influence clinical evaluation. Therefore, we proposed a refined protocol by increasing the number of patient body mass index (BMI) categories from three to seven groups and verified its usefulness by comparing the image sensitivity indicators (S-values) and entrance surface doses (ESDs) of the conventional protocol with those of our refined protocol.

METHODS: This retrospective observational study included 388 datasets (standing position, 224; supine position, 164) for the conventional protocol (December 2019-April 2021) and 336 datasets (standing position, 233; supine position, 103) for the refined protocol (June-November 2021). The conventional protocol (BMI-3 protocol) divided the patients into three BMI groups (BMI < 17, 17≤BMI < 25, and BMI ≥ 25 kg/m2 ), whereas the refined protocol (BMI-7 protocol) divided the patients into seven BMI groups (BMI < 17, 17 ≤ BMI < 20, 20 ≤ BMI < 23, 23 ≤ BMI < 26, 26 ≤ BMI < 29, 29 ≤ BMI < 32, and BMI ≥ 32 kg/m2 ). The coefficients of variation (CVs) for the S-values and ESDs acquired using the two protocols were compared.

RESULTS: The CVs of the S-values in the BMI-7 protocol group were significantly lower than those in the BMI-3 protocol group for the standing (28.8% vs. 16.7%; p < 0.01) and supine (24.5% vs. 17.7%; p < 0.01) positions. The ESDs of patients scanned using the BMI-7 protocol were significantly lower than those scanned using the BMI-3 protocol in the standing (1.3 vs. 1.1 mGy; p < 0.01) and supine positions (2.5 vs. 1.6 mGy; p < 0.01), although the mean BMI of the two groups were similar.

CONCLUSION: We introduced the BMI-7 protocol and demonstrated its standardized image quality and reduced radiation exposure in patients undergoing DCR.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app