Management of a broken guiding catheter tip: Cut and fix technique.

A 49-year-old male presented with class III exertional angina, 1 year after angioplasty of the left anterior descending artery (LAD) and right coronary artery. Coronary angiogram revealed 90% in-stent restenosis (ISR) in mid-LAD with angiographic impression of stent fracture. Optical coherence tomographic evaluation of mid-LAD ISR showed a distinct 3 mm long "eclipse sign" indicating embolized, broken guiding catheter tip as a cause of ISR, which was confirmed on reviewing 1-year-old angiographic images. This was managed with "cut and fix technique" using cutting balloon and another drug-eluting stent. Optical coherence tomographic at 9 months showed well endothelialized stent with a thin layer of neo-intimal hyperplasia over the sandwiched broken guiding tip.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app