Influence of neonatal endotracheal tube dimensions on oscillometry-acquired reactance: a bench study.

OBJECTIVE: To examine the influence of the endotracheal tube (ETT) on respiratory reactance (Xrs) measured with the forced oscillation technique (FOT) and develop a correction method for it.

APPROACH: In a bench study, the reactance of ETTs (Xtube) with different dimensions was measured on a breathing test lung in various respiratory settings.

MAIN RESULTS: Xtube can be accurately predicted by a fitted formula, with an R2 of 0.97, with negligible effects due to changes in respiratory pattern and lung volume.

SIGNIFICANCE: The developed formula offers the ability to measure ETT-independent Xrs values of patients, improving the potential of FOT for lung function testing in mechanically ventilated newborns.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app