[Comparison of point of care capillary international normalized ratio (INR) with INR measured in venous samples].
Revista Médica de Chile 2019 October
BACKGROUND: INR is used to monitor the treatment with vitamin K antagonists. A strategy to reduce waiting times for sampling is to measure INR in a capillary sample using a portable point of care (POC) type coagulometer.
AIM: To evaluate the correlation of CoaguChek Pro II™, Xprecia™ and microINR™ with venous INR measured at the clinical laboratory and their ease of use.
MATERIALS AND METHODS: Patients provided capillary and venous blood samples for parallel tests comparing Xprecia™ Stride with CoaguChek Pro II™ and with venous INR, microINR™ with CoaguChek Pro IITM and with venous INR. The devices' ease of use was assessed surveying the sampling staff.
RESULTS: The three tested devices had good correlation coefficients with venous INR: CoaguChek Pro IITM 0.953 and 0.962; Xprecia™ of 0.912 and microINR™ of 0.932. The correlation coefficient of Xprecia™ with CoaguChek Pro IITM was 0.937 and microINR™ with CoaguChek Pro IITM was 0.976.
CONCLUSIONS: CoaguChek Pro IITM, Xprecia™ and microINR™ results had a good correlation coefficient with INR measured at the laboratory. Our results indicate that, in the hands of trained users, POC-type coagulometers are reliable and acceptable for routine use in anticoagulant treatment control.
AIM: To evaluate the correlation of CoaguChek Pro II™, Xprecia™ and microINR™ with venous INR measured at the clinical laboratory and their ease of use.
MATERIALS AND METHODS: Patients provided capillary and venous blood samples for parallel tests comparing Xprecia™ Stride with CoaguChek Pro II™ and with venous INR, microINR™ with CoaguChek Pro IITM and with venous INR. The devices' ease of use was assessed surveying the sampling staff.
RESULTS: The three tested devices had good correlation coefficients with venous INR: CoaguChek Pro IITM 0.953 and 0.962; Xprecia™ of 0.912 and microINR™ of 0.932. The correlation coefficient of Xprecia™ with CoaguChek Pro IITM was 0.937 and microINR™ with CoaguChek Pro IITM was 0.976.
CONCLUSIONS: CoaguChek Pro IITM, Xprecia™ and microINR™ results had a good correlation coefficient with INR measured at the laboratory. Our results indicate that, in the hands of trained users, POC-type coagulometers are reliable and acceptable for routine use in anticoagulant treatment control.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app