Add like
Add dislike
Add to saved papers

Late failure of a Nellix endoprosthesis treated with the t-Branch off-the-shelf multibranched stent graft.

A 66-year-old man arrived at our emergency department 3 years after an endovascular aneurysm sealing procedure performed at another center. Computed tomography angiography showed distal migration of a Nellix endoprosthesis (Endologix, Irvine, Calif) and a posterior contained rupture. The left kidney was excluded by an occluded iliac-left renal bypass, which was performed at the time of the Nellix implantation because of unintended coverage. A t-Branch (Cook Medical, Bloomington, Ind) was implanted in an emergency, but the delivery caused disruption of the Nellix limb, requiring relining. Target vessels were bridged with VBX stents (W. L. Gore & Associates, Flagstaff, Ariz). The postoperative course was uneventful except for renal function impairment that was restored 2 weeks later.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app