A novel very simple laparoscopic hepatic inflow occlusion apparatus for laparoscopic liver surgery.

Surgical Endoscopy 2019 January
BACKGROUND: Control of bleeding is extremely important for laparoscopic liver resection. We introduce a new and operationally simple laparoscopic hepatic inflow occlusion apparatus (LHIOA) and its successful application in laparoscopic surgery for patients with cirrhosis.

METHODS: The self-designed LHIOA was constructed using a tracheal catheter (7.5#) and infusion set. The tracheal catheter and infusion set were trimmed to 30 and 70 cm, to serve as an occlusion tube and occlusion tape, respectively. After establishment of pneumoperitoneum, the occlusion tape was inserted to encircle the hepatoduodenal ligament. The occlusion tube was then introduced and the ends of the occlusion tape were pulled out of it to occlude the hepatic inflow. Under intermittent vascular occlusion with the LHIOA, the liver parenchyma was transected using an ultrasonic scalpel and monopolar electrocoagulation. Outcomes of the application of the LHIOA in hepatocellular carcinoma patients with cirrhosis (LHIOA group, n = 46) were compared with patients undergoing laparoscopic hepatectomy without LHIOA (non-LHIOA group, n = 46), using one-to-one propensity case-matched analysis.

RESULTS: The LHIOA effectively occluded the hepatic inflow while showing no damage to the hepatoduodenal ligament. The time required for presetting the LHIOA is 6.8 ± 0.6 min. The conversion rate in the non-LHIOA group was 13.0% while there was no conversion in the occlusion group (P < 0.001). The median blood loss of patients in the LHIOA group (60 ml, range 50-200 ml) was significantly less than that of patients in the non-LHIOA group (250 ml, range 100-800) (P < 0.001). Transfusion was required in 8 patients in the non-LHIOA group while no transfusion was required in the LHIOA group. The median operative time in the LHIOA group (157 min, range 80-217 min) was significantly shorter than that in the non-LHIOA group (204 min, range 105-278 min) (P < 0.001).

CONCLUSIONS: The new LHIOA is effective, safe, and simple. It can significantly reduce conversion rate, blood loss, and operative time. It facilitates laparoscopic liver resection and is recommended for use.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app