The mean photon energy ĒF at the point of measurement determines the detector-specific radiation quality correction factor kQ,M in (192)Ir brachytherapy dosimetry.
Zeitschrift Für Medizinische Physik 2016 September
The application of various radiation detectors for brachytherapy dosimetry has motivated this study of the energy dependence of radiation quality correction factor kQ,M, the quotient of the detector responses under calibration conditions at a (60)Co unit and under the given non-reference conditions at the point of measurement, M, occurring in photon brachytherapy. The investigated detectors comprise TLD, radiochromic film, ESR, Si diode, plastic scintillator and diamond crystal detectors as well as ionization chambers of various sizes, whose measured response-energy relationships, taken from the literature, served as input data. Brachytherapy photon fields were Monte-Carlo simulated for an ideal isotropic (192)Ir point source, a model spherical (192)Ir source with steel encapsulation and a commercial HDR GammaMed Plus source. The radial source distance was varied within cylindrical water phantoms with outer radii ranging from 10 to 30cm and heights from 20 to 60cm. By application of this semiempirical method - originally developed for teletherapy dosimetry - it has been shown that factor kQ,M is closely correlated with a single variable, the fluence-weighted mean photon energy ĒF at the point of measurement. The radial profiles of ĒF obtained with either the commercial (192)Ir source or the two simplified source variants show little variation. The observed correlations between parameters kQ,M and ĒF are represented by fitting formulae for all investigated detectors, and further variation of the detector type is foreseen. The herewith established close correlation of radiation quality correction factor kQ,M with local mean photon energy ĒF can be regarded as a simple regularity, facilitating the practical application of correction factor kQ,M for in-phantom dosimetry around (192)Ir brachytherapy sources. ĒF values can be assessed by Monte Carlo simulation or measurement. A technique describing the local measurement of ĒF will be published separately.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app