Understanding coronary artery bypass transit time flow curves: role of bypass graft compliance.

OBJECTIVES: Low mean bypass graft flow (Q) and high pulsatility index (PI) measured by the transit time flow measurement method are not specific for anastomotic stenosis, but occur with competitive flow and poor coronary run-off. We hypothesized that graft compliance is responsible for these changes and that flow measured at the proximal end of the coronary bypass can be viewed as a sum of the graft capacitive flow and flow that passes through the distal anastomosis.

METHODS: Transit time flow measurements (TTFMs) of 15 left internal thoracic artery (LITA) to LAD bypass grafts and 10 saphenous vein grafts (SVGs) to either the right coronary artery (RCA) or posterior descending artery (PDA) were analysed. The TTFM was performed on the proximal and distal end of the graft, and proximally with distal occlusion of the graft. Low mean bypass graft flow PI and diastolic filling (DF) measured distally and proximally were compared, and graft compliance was estimated.

RESULTS: Diastolic filling was higher distally in every single case (LITA-LAD: distal DF 76 ± 12% vs proximal 66 ± 13%, P = 0.005; SVG-RCA/PDA: distal 72 ± 15% vs proximal 63 ± 12%, P = 0.018). There were no significant differences in Q and PI. Subtracting the distal from the proximal flow gave a result identical to the proximal TTFM in distally occluded grafts, confirming the presence of graft capacitive flow. Graft compliance estimated from the flow of distally occluded grafts was 0.99 ± 0.47 μl/mmHg for LITA grafts and 0.78 ± 0.42 μl/mmHg for SVG grafts.

CONCLUSIONS: The study confirmed that the TTFM measured at the proximal end of the coronary bypass could be viewed as a sum of graft capacitive flow and the flow that passes through the distal anastomosis. Graft capacitive flow increases the systolic and decreases the diastolic TTFM when measured at the proximal end of the graft. It explains the higher DF when the TTFM is measured at the distal end of the graft and the increase in the PI at the proximal end when Q decreases. As the influence of graft capacitive flow on the PI in low Q can be eliminated by performing the TTFM at the distal end of the graft, we believe that the value of PI is clinically irrelevant.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app