Add like
Add dislike
Add to saved papers

Patent arterial duct occlusion with two Amplatzer Duct Occluder devices.

It is accepted practice to close large patent arterial ducts (PDA) with Amplatzer duct occluder devices, with extremely low rates of residual PDA. We report a child who required device closure of PDA with two Amplatzer PDA devices on two separate occasions, despite the first device deployment being a standard placement of an appropriately sized Amplatzer device in the usual position.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app