Progressive decrease in amplitude of intracardiac ventricular electrogram and higher left ventricular ejection fraction are associated with conductors' externalization in Riata leads.
AIMS: Increased rates of structural abnormalities including externalized conductors have been reported in the Riata family of implantable cardioverter-defibrillator leads (St Jude Medical). Little is known about their reliability and the time lag for emergence of functional or structural abnormalities.
METHODS AND RESULTS: Thirty-six patients who received small-caliber leads of the Riata family and who completed face-profile flouroscopies, repeated at every 6 months were included. We assessed the prevalence of conductors' externalization and its relation to abnormal electrical parameters or adverse events. Thirty-six patients, mean age = 64 ± 10 years, with at least 7-month completed fluoroscopy follow-up were included in the analysis. Externalized conductors were identified in 12 (33%) patients after a 53-month (13-114) mean delay. A higher left ventricular ejection fraction (LVEF): 47 ± 13 vs. 33 ± 12%, P = 0.04, and a progressive decrease (≥30% of the initial value) in amplitude of ventricular electrogram 9/12 (75%) vs. 4/24 (17%), P = 0.03 were independently associated with the fluoroscopic failures. Detection of the conductors' externalization was preceded by an electrical lead abnormality in 10 (83%) patients.
CONCLUSION: Repeated face-profile fluoroscopies allowed detection of conductors' externalization in 33% of patients implanted with Riata leads. Better LVEF and a progressive decrease in amplitude of intracardiac ventricular electrogram were independently associated with externalized conductors. The structural abnormality was preceded by an electrical lead dysfunction 83% of patients.
METHODS AND RESULTS: Thirty-six patients who received small-caliber leads of the Riata family and who completed face-profile flouroscopies, repeated at every 6 months were included. We assessed the prevalence of conductors' externalization and its relation to abnormal electrical parameters or adverse events. Thirty-six patients, mean age = 64 ± 10 years, with at least 7-month completed fluoroscopy follow-up were included in the analysis. Externalized conductors were identified in 12 (33%) patients after a 53-month (13-114) mean delay. A higher left ventricular ejection fraction (LVEF): 47 ± 13 vs. 33 ± 12%, P = 0.04, and a progressive decrease (≥30% of the initial value) in amplitude of ventricular electrogram 9/12 (75%) vs. 4/24 (17%), P = 0.03 were independently associated with the fluoroscopic failures. Detection of the conductors' externalization was preceded by an electrical lead abnormality in 10 (83%) patients.
CONCLUSION: Repeated face-profile fluoroscopies allowed detection of conductors' externalization in 33% of patients implanted with Riata leads. Better LVEF and a progressive decrease in amplitude of intracardiac ventricular electrogram were independently associated with externalized conductors. The structural abnormality was preceded by an electrical lead dysfunction 83% of patients.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app