Constrained cups appear incapable of meeting the demands of revision THA.

BACKGROUND: Failure rates of constrained cups for treating recurrent dislocation in revision THA range from 40% to 100%. Although constrained liners are intended to stabilize the hip by mechanically preventing dislocation, the resulting loss of range of motion may lead to impingement and, ultimately, implant failure.

QUESTIONS/PURPOSES: We therefore documented the mechanisms of failure of constrained acetabular cups in revision THA and determined the type and severity of damage (wear, fracture, and impingement) that occurs in situ.

METHODS: We retrieved 57 constrained components of four different designs at revision THA and examined for the presence of rim impingement, oxidation, cracks within the liner, backside wear, pitting, scratching, abrasion, burnishing, and the presence of embedded particles. Articular wear was calculated from the volume of the concave articular bearing surface, which was measured using the fluid displacement method.

RESULTS: Failure of the locking ring was responsible for 51% of failures, whereas 28% of revisions were the result of acetabular cup loosening, 6% backside wear, and 22% infection. Impingement damage of the rim of the polyethylene liner was seen in all retrievals with moderate or severe damage in 54%. The average volumetric wear rate of the articular surface was 95 mm(3)/year.

CONCLUSIONS: Failure of the locking liner ring and loosening of the acetabular cup are the primary causes of mechanical failure with constrained liners; polyethylene is an inadequate material for restricting motion of the hip to prevent instability. The durability of these devices is unlikely to improve unless the mechanical demands are modified through increased range of motion leading to less frequent rim impingement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app