Case Reports
Journal Article
Add like
Add dislike
Add to saved papers

Abdominal aortic embolization of a Figulla atrial septum occluder device, at the level of the celiac axis, after an atrial septal defect closure: hybrid attempt.

Vascular 2010 January
A 41-year-old woman was treated with a Figulla (Occlutec, Helsingborg, Sweden) atrial septum occluder device with no intraprocedural complications. Five months later, dislocation of the device in the abdominal aorta was detected. The occluder device was located at the level of the celiac axis, nearly obstructing the entire aorta. Owing to total incorporation of the device, endoluminal retrieval was not possible. Through a medial rotation approach, the device was safely removed. This is a rare complication after endoluminal closure of an atrial septum defect. The retrieval possibilities are discussed.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app