Comparative Study
Journal Article
Add like
Add dislike
Add to saved papers

The risk of notching the anterior femoral cortex with the use of navigation systems in total knee arthroplasty.

Use of navigation systems has recently been introduced in total knee arthroplasty (TKA) to achieve more reliable prosthetic alignment. In the sagittal plane, there are two important requirements for navigation systems: (1) perpendicular cut to the femoral mechanical axis and (2) prevention of notching of anterior femoral cortex. These two requirements, however, may conflict. The angles between the line of the anterior femoral cortex and four sagittal femoral mechanical axes for navigation systems using radiographs of the entire lower extremity, while standing were measured and compared. These four sagittal axes simulated on the radiographs in navigation systems were in extension relative to the line of the anterior femoral cortex in 40-85% of cases in male and 65-100% in elderly female. The present study showed that navigation systems have the potential risk for notching of anterior femoral cortex.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app