Utilization of shape memory in external vein-graft meshes allows extreme diameter constriction for suppressing intimal hyperplasia: a non-human primate study.

OBJECTIVE: Constrictive external Nitinol meshes have been shown to suppress neointimal tissue formation and preserve endothelial integrity in vein grafts. As this mitigating effect increased with the degree of constriction, we investigated whether extreme constriction was possible without leading to detrimental luminal encroachment.

METHODS: A senescent non-human primate model (Chacma baboons/bilateral femoral interposition grafts) mimicking the clinical size-mismatch between vein grafts and run-off arteries was used. Control grafts were either untreated (group 1) or spray-coated with fibrin glue (group 2). Nitinol meshes constricting the lumen by 90% (group 4). Anastomotic size mismatch at implantation was expressed as quotient of cross-sectional area of run-off artery to vein graft (Q(C)).

RESULTS: At 6 months, all vein grafts without mesh support showed thick eccentric layers of neointimal tissue (group 1: 348 +/- 130 microm [Q(C) median at implant 0.19]; group 2: 318 +/- 142 microm [Q(C) median at implant 0.17]). Fibrin glue-spraying had no effect. In contrast, neointimal tissue was absent in all mesh-supported grafts (P < .007 in all cases) both at 6 weeks/6 months (group 3: 7.5 +/- 8.8 mum and 2.5 +/- 4.4 microm [Q(C) median at implant 1.47]; group 4: 1.3 +/- 0.6 microm and 3.8 +/- 5.6 microm [Q(C) median at implant 3.09]). Except for mild tissue buckling (fold height <356 microm) in one group 3 graft, none of the mesh-constricted grafts showed wall folds. Endothelial coverage was only complete in the mesh-supported groups (100% in group 3 and 4 vs 85 +/- 14%; P < .023 in group 1). Fibrin glue alone (52 +/- 48%) did not preserve endothelialization of control grafts (P < .38).

CONCLUSION: Extreme vein graft constriction using external Nitinol meshes is possible without detrimental tissue buckling. Although moderate constriction was found to be sufficient for mitigating diffuse intimal hyperplasia and endothelial detachment, extreme constriction may occasionally be required to eliminate luminal irregularities.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app