COMPARATIVE STUDY
JOURNAL ARTICLE

Mechanical and handling properties of braided polyblend polyethylene sutures in comparison to braided polyester and monofilament polydioxanone sutures

Daniel M Wüst, Dominik C Meyer, Philippe Favre, Christian Gerber
Arthroscopy 2006, 22 (11): 1146-53
17084288

PURPOSE: This study was designed to comprehensively compare the mechanical properties of 4 types of braided polyblend sutures with widely used braided polyester and monofilament polydioxanone sutures.

METHODS: Polyblend polyethylene sutures (FiberWire [Arthrex, Naples, FL], Herculine [Linvatec, Largo, FL], Orthocord [DePuy Mitek, Raynham, MA], and Ultrabraid [Smith & Nephew Endoscopy, Andover, MA]), a braided polyester suture (Ethibond; Ethicon, Somerville, NJ), and an absorbable monofilament polydioxanone suture (PDS II; Ethicon), all USP No. 2, were mechanically tested. Fraying resistance was tested on eyelets of metallic and absorbable suture anchors. Cartilage abrasion caused by an intra-articularly placed suture knot was simulated by fraying on distal porcine femora.

RESULTS: All polyblend sutures were stronger than Ethibond or PDS II sutures by at least a factor of 2, with or without a knot. When knotted, Herculine (261 +/- 44 N) was strongest, followed by Ultrabraid (244 +/- 3 N). FiberWire was most resistant against fraying on metallic anchors. Orthocord was by far least abrasive with absorbable anchors. Resistance to fraying was 100- to 500-fold (absorbable anchors) and 6- to 30-fold (metallic anchor) better for all polyblend sutures than for Ethibond. All braided sutures caused a similar amount of abrasion of joint cartilage, but they caused significantly more abrasion (>20-fold) than the monofilament degradable suture.

CONCLUSIONS: The ultimate strength of polyblend suture material was 2- to 2.5-fold greater than that of polyester or polydioxanone sutures, but the resistance to fraying was up to 500-fold greater than that of polyester or polydioxanone sutures. With regard to strength, this makes polyblend sutures particularly advantageous for use with metallic edges of anchors or prostheses or with absorbable anchor eyelets.

CLINICAL RELEVANCE: With a high resistance to fraying against metallic edges or a decrease in cutting of absorbable suture eyelets being up to 500-fold greater than with polyester or polydioxanone sutures, the new polyblend sutures appear to fill a void in the armamentarium of the surgeon, provided that at least 2 throws more than with conventional sutures are used for knot tying.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
17084288
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"