We have located links that may give you full text access.
Clinical Trial
Journal Article
Randomized Controlled Trial
Video networking of cardiac catheterization laboratories.
American Heart Journal 1999 Februrary
BACKGROUND: The purpose of this study was to assess the feasibility and accuracy of a video telecommunication network to transmit coronary images to provide on-line interaction between personnel in a cardiac catheterization laboratory and a remote core laboratory.
METHODS: A telecommunication system was installed in the cardiac catheterization laboratory at Kaiser Hospital, Los Angeles, and the core laboratory at the University of California, Irvine, approximately 40 miles away. Cineangiograms, live fluoroscopy, intravascular ultrasound studies and images of the catheterization laboratory were transmitted in real time over a dedicated T1 line at 768 kilobytes/second at 15 frames/second. These cases were performed during a clinical study of angiographic guidance versus intravascular ultrasound (IVUS) guidance of stent deployment. During the cases the core laboratory performed quantitative analysis of the angiograms and ultrasound images. Selected images were then annotated and transmitted back to the catheterization laboratory to facilitate discussion during the procedure.
RESULTS: A successful communication hookup was obtained in 39 (98%) of 40 cases. Measurements of angiographic parameters were very close between the original cinefilm and the transmitted images. Quantitative analysis of the ultrasound images showed no significant difference in any of the diameter or cross-sectional area measurements between the original ultrasound tape and the transmitted images. The telecommunication link during the interventional procedures had a significant impact in 23 (58%) of 40 cases affecting the area to be treated, the size of the inflation balloon, recognition of stent underdeployment, or the existence of disease in other areas that was not noted on the original studies.
CONCLUSIONS: Current video telecommunication systems provide high-quality images on-line with accurate representation of cineangiograms and intravascular ultrasound images. This system had a significant impact on 58% of the cases in this small clinical trial. Telecommunication networks between hospitals and a central core laboratory may facilitate physician training and improve technical skills and judgement during interventional procedures. This project has implications for how multicenter clinical trials could be operated through telecommunication networks to ensure conformity with the protocol.
METHODS: A telecommunication system was installed in the cardiac catheterization laboratory at Kaiser Hospital, Los Angeles, and the core laboratory at the University of California, Irvine, approximately 40 miles away. Cineangiograms, live fluoroscopy, intravascular ultrasound studies and images of the catheterization laboratory were transmitted in real time over a dedicated T1 line at 768 kilobytes/second at 15 frames/second. These cases were performed during a clinical study of angiographic guidance versus intravascular ultrasound (IVUS) guidance of stent deployment. During the cases the core laboratory performed quantitative analysis of the angiograms and ultrasound images. Selected images were then annotated and transmitted back to the catheterization laboratory to facilitate discussion during the procedure.
RESULTS: A successful communication hookup was obtained in 39 (98%) of 40 cases. Measurements of angiographic parameters were very close between the original cinefilm and the transmitted images. Quantitative analysis of the ultrasound images showed no significant difference in any of the diameter or cross-sectional area measurements between the original ultrasound tape and the transmitted images. The telecommunication link during the interventional procedures had a significant impact in 23 (58%) of 40 cases affecting the area to be treated, the size of the inflation balloon, recognition of stent underdeployment, or the existence of disease in other areas that was not noted on the original studies.
CONCLUSIONS: Current video telecommunication systems provide high-quality images on-line with accurate representation of cineangiograms and intravascular ultrasound images. This system had a significant impact on 58% of the cases in this small clinical trial. Telecommunication networks between hospitals and a central core laboratory may facilitate physician training and improve technical skills and judgement during interventional procedures. This project has implications for how multicenter clinical trials could be operated through telecommunication networks to ensure conformity with the protocol.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app