Clinical experience with 118 brain tissue oxygen partial pressure catheter probes.

Neurosurgery 1998 November
OBJECTIVE: We assessed the technical and diagnostic reliability of partial pressure of oxygen (PO2) of brain tissue (P(ti)O2) monitoring. The monitoring system and the catheter probes were tested in vitro, and clinical experiences obtained with 118 brain P(ti)O2 catheter probes, used in 101 patients, are reported.

METHODS: The polarographic (LICOX; Medical Systems Corp., Greenvale, NY) P(ti)O2 catheter probe lies 22 to 27 mm below the dura level; its PO2-sensitive surface is 7.1 mm2. For 10 patients, the adaptation time (with initially unreliable signals after insertion) was determined. For 27 patients, the probe was removed in a stepwise fashion (three increments of 5 mm) and the heterogeneity of brain P(ti)O2 levels was investigated. After removal of the catheter probes, their PO2 and zero display error values were determined and compared with probe performance data obtained in vitro with unused PO2 catheter probes.

RESULTS: Small iatrogenic hematomas were observed for two patients (1.7%). No infection occurred after 6.7 +/- 3.9 days (mean +/- standard deviation) of monitoring. The technical complication (dislocation or defect) rate was 13.6%. The mean adaptation time was 79.0 +/- 51.7 min. A flow chart is presented, which helps to rule out artifacts. The mean P(ti)O2 measured at 22 to 27 mm below the dura was 23.8 +/- 8.1 mm Hg, at 17 to 22 mm was 25.7 +/- 8.3 mm Hg, at 12 to 17 mm was 33.0 +/- 13.3 mm Hg (P < 0.01, compared with the initial value), and at 7 to 12 mm was 33.3 +/- 13.3 mm Hg (P < 0.01). Recent catheter probe versions exhibited a PO2 display error of -1.2 +/- 5.1% (mean +/- standard deviation, n = 38) and a mean zero display error of 1.1 +/- 0.9 mm Hg (n = 34). The greatest PO2 display errors were measured during the first 4 days of continuous monitoring. In the in vitro test (of 12 unused catheter probes), the maximal probe display error was 1.07 +/- 2.14%, tested at temperatures between 22 degrees C and 37 degrees C and tested at oxygen pressures of 0, 44, and 150 mm Hg. In vitro, the zero display error was -0.21 +/- 0.25 mm Hg.

CONCLUSION: Brain P(ti)O2 monitoring, reflecting an area 17 to 27 mm below the dura, is a safe and reliable technique for monitoring cerebral oxygenation. Excluding the first 1 hour after insertion, data are reliable, with almost 100% good data quality.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

Managing Alcohol Withdrawal Syndrome.Annals of Emergency Medicine 2024 March 26

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app