Comparison of exercise performance in patients with chronic severe heart failure versus left ventricular assist devices.

Circulation 1998 September 23
BACKGROUND: Left ventricular assist devices (LVADs) are frequently used as a bridge to cardiac transplantation and may be useful as long-term therapy. The purpose of this study was to compare the exercise performance of LVAD patients with that of ambulatory heart failure patients.

METHODS AND RESULTS: Exercise testing with hemodynamic and respiratory gas measurements was performed in 65 congestive heart failure (CHF) patients (age 53+/-10 years) and 20 LVAD patients (age 49+/-8 years). Peak Vo2 was significantly higher in the LVAD than the CHF patients (CHF, 12+/-3; LVAD, 15. 9+/-3.8 mL . kg-1 . min-1; P<0.001), as was the Vo2 at the anaerobic threshold (CHF, 8.1+/-2.1; LVAD, 12.2+/-2.9 mL . kg-1 . min-1; P<0.001). At rest, mean arterial blood pressure (CHF, 87+/-11; LVAD, 94+/-9 mm Hg) and cardiac output (CHF, 4+/-1; LVAD, 4. 9+/-0.9 L/min) were increased, whereas mean pulmonary artery pressure (CHF, 28+/-11; LVAD, 18+/-4 mm Hg) and pulmonary artery wedge pressure (CHF, 16+/-10; LVAD 5+/-3 mm Hg) were reduced (all P<0.01). At peak exercise, heart rate (CHF,125+/-24; LVAD, 148+/-24 bpm), blood pressure (CHF, 87+/-14; LVAD,96+/-12 mm Hg), and cardiac output (CHF, 7.6+/-2.2; LVAD, 11.2+/-2.6 L/min) were higher (all P<0. 01), whereas mean pulmonary artery pressure (CHF, 48+/-12; LVAD, 30+/-5 mm Hg) and mean pulmonary capillary wedge pressure (CHF, 31+/-11; LVAD, 14+/-6 mm Hg) were lower in the LVAD group (both P<0. 001). In the LVAD patients, Fick cardiac output was higher than LVAD flow sensor value measurements (Fick, 11.6+/-2.5; LVAD, 8.1+/-1.2 L/min; P<0.001).

CONCLUSIONS: Hemodynamic measurements at rest and during exercise are significantly improved in patients with devices compared with those in ambulatory heart failure patients awaiting cardiac transplantation. Similarly, the exercise capacity of device patients is better than that of transplant candidates and in the majority of patients is similar to that of patients with mild CHF.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app