OBJECTIVE: The biomechanical stability of supracondylar femoral fractures fixed with a condylar blade plate (plate), a Green Seligson Henry nail (GSHN), or a new retrograde unreamed supracondylar femoral nail (new nail) based on the AO unreamed femoral nail were compared.
DESIGN: A standardized simulated comminuted supracondylar femoral fracture (segmental defect) in fresh frozen paired cadaveric femora was stabilized with one of the implants. The interfragmentary fracture site stiffness in three directions and axial strength of the fixator-bone construct were compared (pairwise).
RESULTS: The plate versus the new nail was (a) axially 10 percent as stiff and 50 percent as strong (ultimate strength), (b) as stiff in A/P bending, and (c) five times more stiff in torsion. Varus angle at failure under axial load was significantly greater for the plate than for the new nail. There were no statistical differences in axial stiffness and ultimate strength between the new nail and the GSHN, but the new nail was 50 percent and 30 percent as stiff in A/P bending and torsion, respectively. The magnitude of deformation at failure under axial loading was similar.
CONCLUSIONS: In fixation of extraarticular comminuted supracondylar distal femur fractures, results indicate that (a) the new nail provides equal or greater stability than does the plate, except when large torsional loads are anticipated, and (b) the new nail provides stability equal to the GSHN for axial loading and lesser stability against off-axis loads. As is evident in this and other studies, intramedullary implants are less torsionally stiff than are plates. The torsional stiffness of the new nail is expected to be sufficient because it is comparable to many available nails, and low torsional moments are expected for healing femoral supracondylar fractures.
DESIGN: A standardized simulated comminuted supracondylar femoral fracture (segmental defect) in fresh frozen paired cadaveric femora was stabilized with one of the implants. The interfragmentary fracture site stiffness in three directions and axial strength of the fixator-bone construct were compared (pairwise).
RESULTS: The plate versus the new nail was (a) axially 10 percent as stiff and 50 percent as strong (ultimate strength), (b) as stiff in A/P bending, and (c) five times more stiff in torsion. Varus angle at failure under axial load was significantly greater for the plate than for the new nail. There were no statistical differences in axial stiffness and ultimate strength between the new nail and the GSHN, but the new nail was 50 percent and 30 percent as stiff in A/P bending and torsion, respectively. The magnitude of deformation at failure under axial loading was similar.
CONCLUSIONS: In fixation of extraarticular comminuted supracondylar distal femur fractures, results indicate that (a) the new nail provides equal or greater stability than does the plate, except when large torsional loads are anticipated, and (b) the new nail provides stability equal to the GSHN for axial loading and lesser stability against off-axis loads. As is evident in this and other studies, intramedullary implants are less torsionally stiff than are plates. The torsional stiffness of the new nail is expected to be sufficient because it is comparable to many available nails, and low torsional moments are expected for healing femoral supracondylar fractures.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app