CLINICAL TRIAL
COMPARATIVE STUDY
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Mainstream end-tidal carbon dioxide monitoring in the neonatal intensive care unit.

Pediatrics 1998 April
BACKGROUND: Continuous noninvasive monitoring of arterial carbon dioxide (CO2) in neonatal intensive care unit (NICU) patients would help clinicians avoid complications of hypocarbia and hypercarbia. End-tidal CO2 monitoring has not been used in this population to date, but recent technical advances and the introduction of surfactant therapy, which improves ventilation-perfusion matching, might improve the clinical utility of end-tidal monitoring.

OBJECTIVE: To determine the accuracy and precision of end-tidal CO2 monitoring in NICU patients.

DESIGN: Nonrandomized recording of simultaneous end-tidal and arterial CO2 pairs.

SETTING: Two university NICUs.

PATIENTS: Forty-five newborn infants receiving mechanical ventilation who had indwelling arterial access, and a predefined subsample of infants who were <1000 g birth weight, <8 days of age, and who received surfactant therapy (extremely low birth weight -ELBW- <8).

OUTCOME MEASURES: The correlation coefficient, degree of bias, and 95% confidence interval were determined for both the overall population and the ELBW <8 subgroup. Those factors which significantly influenced the bias were identified. The ability of the end-tidal monitor to alert the clinician to instances of hypocarbia or hypercarbia was determined.

RESULTS: There were 411 end-tidal/arterial pairs analyzed from 45 patients. The correlation coefficient was 0.833 and the bias was -6. 9 mm Hg (95% confidence interval, +/-11.5 mm Hg). The results did not differ markedly in the ELBW <8 infants. Measures of the degree of lung disease, the ventilation index and the oxygenation index, had small influences on the degree of bias. This type of capnometry identified 91% of the instances when the arterial CO2 pressure was between 34 and 54 mm Hg using an end-tidal range of 29 to 45 mm Hg. End-tidal values outside this range had a 63% accuracy in predicting hypocarbia or hypercarbia.

CONCLUSION: End-tidal CO2 monitoring in NICU patients is as accurate as capillary or transcutaneous monitoring but less precise than the latter. It may be useful for trending or for screening patients for abnormal arterial CO2 values.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app