Single lead VDD pacing in children with complete heart block.
Canadian Journal of Cardiology 1998 January
OBJECTIVE: To evaluate clinically transvenous single lead VDD pacing in children with complete heart block.
DESIGN: Case series.
SETTING: Tertiary care unit of a pediatric teaching hospital.
PATIENTS: Consecutive sample of children with complete heart block requiring pacing and meeting the criteria for VDD pacing.
INTERVENTIONS: VDD single lead pacing systems (Thera VDD 8948 generator, Capsure VDD 5032 single pass lead) were implanted in nine children (mean age 10.7 years, 95% CI 8.9 to 12.3). All patients were reviewed six weeks and six months post-implantation with clinical review, detailed pacemaker interrogation, optimal device programming, treadmill exercise testing, 24 h Holter monitoring and echocardiography.
RESULTS: Mean procedure time was 62.5 mins (CI 52 to 73), mean fluoroscopy time 3.8 mins (CI 2.4 to 5.1) and mean P wave sensing amplitude 2.75 mV (CI 2.5 to 3), without significant decrease during Valsalva manoeuvre (mean 2.4 mV, CI 2.2 to 2.8). Low pacing thresholds of mean amplitude 0.39 V (CI 0.38 to 0.4) were established at pulse width 0.5 ms. Two patients required early revision--one lead dislodged and one patient had a loose set screw. Follow-up showed satisfactory pacemaker function, and mean percentage atrioventricular (AV) synchronous pacing 96.8% (CI 93 to 99) at six weeks and 97% (CI 95 to 99) at six months. P wave sensing failures were less than 1%. AV synchronous pacing was maintained for more than 99% of the period of treadmill exercise testing. There was a significant mean increase in cardiac output of 24% (CI 8 to 40, P < 0.05) at rest in VDD mode compared with VVIR mode.
CONCLUSIONS: Single lead VDD pacing is an effective and reliable method of 'physiological' rate responsive AV synchronous pacing in children with complete heart block and should be considered a viable alternative to dual lead pacing systems.
DESIGN: Case series.
SETTING: Tertiary care unit of a pediatric teaching hospital.
PATIENTS: Consecutive sample of children with complete heart block requiring pacing and meeting the criteria for VDD pacing.
INTERVENTIONS: VDD single lead pacing systems (Thera VDD 8948 generator, Capsure VDD 5032 single pass lead) were implanted in nine children (mean age 10.7 years, 95% CI 8.9 to 12.3). All patients were reviewed six weeks and six months post-implantation with clinical review, detailed pacemaker interrogation, optimal device programming, treadmill exercise testing, 24 h Holter monitoring and echocardiography.
RESULTS: Mean procedure time was 62.5 mins (CI 52 to 73), mean fluoroscopy time 3.8 mins (CI 2.4 to 5.1) and mean P wave sensing amplitude 2.75 mV (CI 2.5 to 3), without significant decrease during Valsalva manoeuvre (mean 2.4 mV, CI 2.2 to 2.8). Low pacing thresholds of mean amplitude 0.39 V (CI 0.38 to 0.4) were established at pulse width 0.5 ms. Two patients required early revision--one lead dislodged and one patient had a loose set screw. Follow-up showed satisfactory pacemaker function, and mean percentage atrioventricular (AV) synchronous pacing 96.8% (CI 93 to 99) at six weeks and 97% (CI 95 to 99) at six months. P wave sensing failures were less than 1%. AV synchronous pacing was maintained for more than 99% of the period of treadmill exercise testing. There was a significant mean increase in cardiac output of 24% (CI 8 to 40, P < 0.05) at rest in VDD mode compared with VVIR mode.
CONCLUSIONS: Single lead VDD pacing is an effective and reliable method of 'physiological' rate responsive AV synchronous pacing in children with complete heart block and should be considered a viable alternative to dual lead pacing systems.
Full text links
Related Resources
Trending Papers
Catastrophic Antiphospholipid Syndrome: A Review of Current Evidence and Future Management Practices.Curēus 2024 September
Paroxysmal Nocturnal Hemoglobinuria, Pathophysiology, Diagnostics, and Treatment.Transfusion Medicine and Hemotherapy 2024 October
2024 Guideline for the Primary Prevention of Stroke: A Guideline From the American Heart Association/American Stroke Association.Stroke; a Journal of Cerebral Circulation 2024 October 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app