Monitoring of radio frequency tissue ablation in an interventional magnetic resonance environment. Preliminary ex vivo and in vivo results.
Investigative Radiology 1997 November
RATIONALE AND OBJECTIVES: The authors evaluate the feasibility of monitoring radio frequency (RF) ablation in an interventional, open-configuration, 0.5-tesla magnetic resonance (MR) environment.
METHODS: Ex vivo and in vivo RF coagulation necrosis were induced in porcine paraspinal muscle tissue using a 300 kHz monopolar RF generator applying 5 to 20 W over 3 to 9 minutes. Images were acquired simultaneous to RF application, after RF application, and in an intermittent mode (60 seconds of RF followed by 15 seconds of MR imaging). Temperature changes were monitored based on amplitude (ex vivo) and phase alterations (in vivo) of a T1-weighted graded refocused echo (GRE) sequence enabling an update every 2.5 seconds. A standardized color-coded subtraction technique enhanced signal changes. Additionally, T2- and T1-weighted spin echo (SE) images were acquired with and without intravenous contrast. Macroscopic coagulation size was compared with lesion size seen on MR images.
RESULTS: Although lesion diameters were related directly to applied RF power, the application mode had no significant impact on coagulation size (P > 0.05). As could be expected, MR imaging during RF ablation resulted in major image distortion. Radio frequency effects were seen on images acquired in the continuous and intermittent modes. Coagulation size seen on GRE images correlated well with macroscopy both ex vivo (r = 0.89) and in vivo (r = 0.92). Poorer correlation was found with postinterventional SE sequences (r = 0.78-0.84).
CONCLUSIONS: Magnetic resonance monitoring of RF effects is feasible both ex vivo as well as in vivo using temperature-sensitive sequences in an open-configuration MR environment.
METHODS: Ex vivo and in vivo RF coagulation necrosis were induced in porcine paraspinal muscle tissue using a 300 kHz monopolar RF generator applying 5 to 20 W over 3 to 9 minutes. Images were acquired simultaneous to RF application, after RF application, and in an intermittent mode (60 seconds of RF followed by 15 seconds of MR imaging). Temperature changes were monitored based on amplitude (ex vivo) and phase alterations (in vivo) of a T1-weighted graded refocused echo (GRE) sequence enabling an update every 2.5 seconds. A standardized color-coded subtraction technique enhanced signal changes. Additionally, T2- and T1-weighted spin echo (SE) images were acquired with and without intravenous contrast. Macroscopic coagulation size was compared with lesion size seen on MR images.
RESULTS: Although lesion diameters were related directly to applied RF power, the application mode had no significant impact on coagulation size (P > 0.05). As could be expected, MR imaging during RF ablation resulted in major image distortion. Radio frequency effects were seen on images acquired in the continuous and intermittent modes. Coagulation size seen on GRE images correlated well with macroscopy both ex vivo (r = 0.89) and in vivo (r = 0.92). Poorer correlation was found with postinterventional SE sequences (r = 0.78-0.84).
CONCLUSIONS: Magnetic resonance monitoring of RF effects is feasible both ex vivo as well as in vivo using temperature-sensitive sequences in an open-configuration MR environment.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app