Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.
Critical Care Medicine 1996 October
OBJECTIVE: To evaluate the accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.
DESIGN: Prospective data collection.
SETTING: Emergency department in a 2,000-bed hospital.
PATIENTS: Thirty-nine patients admitted to the emergency department requiring invasive arterial blood pressure monitoring were enrolled to the study protocol.
INTERVENTIONS: Continuous noninvasive blood pressure measurement was performed on the middle phalanx of the second and third finger, using a test instrument which provides continuous arterial waveform display with the use of a finger cuff. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery and direct transduction of the systemic arterial pressure waveform.
MEASUREMENTS AND MAIN RESULTS: Three thousand one hundred eighteen pairs of simultaneous finger cuff and intra-arterial blood pressure measurements were collected in 1-min intervals from 39 patients over a total of 51.8 hrs. The overall discrepancy between both measurements was 0.10 mm Hg. The standard deviation of the differences was +/- 5.02 mm Hg. The mean bias in patients treated with catecholamines was 0.01 mm Hg and was not different from the bias observed in patients without catecholamines (mean bias: 0.23 mm Hg; p > .22). Whereas 95% of all comparisons between finger cuff and intra-arterial measurement had a discrepancy < or = +/- 10 mm Hg, 4.7% had a discrepancy between +/- 10.1 to 15 mm Hg and 0.3% exhibited a discrepancy > +/- 15 mm Hg. In 29 (74%) patients, the duration of errors > 10 mm Hg was < or = 1 min. In seven (18%) patients, the duration of errors > 10 mm Hg was between 2 to 3 mins and in three (8%) patients, the errors lasted for > 3 mins.
CONCLUSIONS: Our data provide a guide to the accuracy and reliability of noninvasive finger blood pressure measurements in critically ill patients. Although most test instrument measurements were reliable, in 8% of all patients large discrepancies (> 10 mm Hg) between both measurements with a duration of > 3 mins were noted. Concerning the considerable risk for arterial cannulation, our preliminary data demonstrate that the test instrument (PORTAPRES, TNO Biomedical Instrumentation Research Unit; The Netherlands) is an advance in noninvasive monitoring of critically ill patients and may be useful in most emergency clinical settings.
DESIGN: Prospective data collection.
SETTING: Emergency department in a 2,000-bed hospital.
PATIENTS: Thirty-nine patients admitted to the emergency department requiring invasive arterial blood pressure monitoring were enrolled to the study protocol.
INTERVENTIONS: Continuous noninvasive blood pressure measurement was performed on the middle phalanx of the second and third finger, using a test instrument which provides continuous arterial waveform display with the use of a finger cuff. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery and direct transduction of the systemic arterial pressure waveform.
MEASUREMENTS AND MAIN RESULTS: Three thousand one hundred eighteen pairs of simultaneous finger cuff and intra-arterial blood pressure measurements were collected in 1-min intervals from 39 patients over a total of 51.8 hrs. The overall discrepancy between both measurements was 0.10 mm Hg. The standard deviation of the differences was +/- 5.02 mm Hg. The mean bias in patients treated with catecholamines was 0.01 mm Hg and was not different from the bias observed in patients without catecholamines (mean bias: 0.23 mm Hg; p > .22). Whereas 95% of all comparisons between finger cuff and intra-arterial measurement had a discrepancy < or = +/- 10 mm Hg, 4.7% had a discrepancy between +/- 10.1 to 15 mm Hg and 0.3% exhibited a discrepancy > +/- 15 mm Hg. In 29 (74%) patients, the duration of errors > 10 mm Hg was < or = 1 min. In seven (18%) patients, the duration of errors > 10 mm Hg was between 2 to 3 mins and in three (8%) patients, the errors lasted for > 3 mins.
CONCLUSIONS: Our data provide a guide to the accuracy and reliability of noninvasive finger blood pressure measurements in critically ill patients. Although most test instrument measurements were reliable, in 8% of all patients large discrepancies (> 10 mm Hg) between both measurements with a duration of > 3 mins were noted. Concerning the considerable risk for arterial cannulation, our preliminary data demonstrate that the test instrument (PORTAPRES, TNO Biomedical Instrumentation Research Unit; The Netherlands) is an advance in noninvasive monitoring of critically ill patients and may be useful in most emergency clinical settings.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app