STUDY OBJECTIVES: To confirm the ability of the esophageal detector device (EDD) to indicate positioning of endotracheal tubes (ETTs) in patients intubated under emergency conditions and to compare the performance of the EDD with that of end-tidal carbon dioxide (ETCO2).

METHODS: This single-subject study comprising a prospective case series was conducted in the emergency department of an urban university hospital. All adult patients were intubated either in the ED or by paramedics in the field. ETT position was initially evaluated by means of auscultation, then EDD, and, finally, spectrographic qualitative ETCO2 monitoring in each patient. Discrepancies between the EDD and ETCO2 results were resolved by means of direct laryngoscopy.

RESULTS: In 100 intubated patients, both the EDD and ETCO2 monitoring detected the single esophageal intubation that occurred. Of the remaining 99 tracheal intubations, the EDD correctly indicated tracheal placement in 98 (sensitivity, 99%) and was indeterminate in 1 case because of blockage of the ETT by secretions resulting from pulmonary edema. By comparison, ETCO2 monitoring correctly indicated tracheal placement in 86 cases (sensitivity, 87%) and was incorrect in 13 cases (P < .01). ETCO2 monitoring failed in 2 patients with pulmonary edema and in 11 patients with cardiac arrest. Among the 37 patients in the cardiac arrest group, the EDD correctly indicated ETT placement in 37 patients (sensitivity, 100%). In contrast, ETCO2 monitoring correctly indicated ETT placement in 26 patients (sensitivity, 70%; P < .01).

CONCLUSION: The EDD reliably confirms tracheal intubation in the emergency patient population. The EDD is more accurate than ETCO2 monitoring in the overall emergency patient population because of its greater accuracy in cardiac arrest patients. [Bozeman WP, Hexter D, Liang HK, Kelen GD: Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app