Evaluation of capnography in nonintubated emergency department patients with respiratory distress.
Academic Emergency Medicine 1995 October
OBJECTIVE: To evaluate the ability of noninvasive capnographic measurement of end-tidal CO2 tension (PetCO2) to predict arterial CO2 tension (PaCO2) in nonintubated ED patients with respiratory distress.
METHODS: A prospective, nonblind study was performed in a level I trauma center/community teaching hospital ED. Participants included all nonintubated adult patients with respiratory distress requiring measurement of arterial blood gases (ABGs); 29 patients were enrolled. PetCO2 was measured with a capnography monitor, using both baseline tidal volumes and forced expiratory volumes. The bias between PetCO2 values and simultaneous measurements of PaCO2 by ABG was assessed.
RESULTS: PetCO2, measured with forced expiration, and PaCO2 agreed well, with bias (i.e., average difference) = 0.44 +/- 0.52 kPa (3.3 +/- 3.9 torr). PetCO2 measured with the tidal volume breath produced an unacceptably high bias of 0.82 +/- 0.70 kPa (6.1 +/- 5.2 torr). Levels of agreement between PaCO2 were similar for smokers and nonsmokers and for men and women. The arterial-end-tidal CO2 tension (Pa-etCO2) difference was not related to PaCO2. Pa-etCO2 correlated with age (r = 0.473; p = 0.01), and was significantly higher in patients with pulmonary disease (1.32 +/- 0.56 kPa; 9.9 +/- 4.2 torr) than it was in those without pulmonary disease (0.46 +/- 0.55 kPa; 3.5 +/- 4.1 torr; p < 0.001).
CONCLUSIONS: Noninvasive PetCO2 monitoring may adequately predict PaCO2 in nonintubated ED patients with respiratory distress who are able to produce a forced expiration. PetCO2 is less accurate for PaCO2 with tidal volume breathing and in patients with pulmonary disease.
METHODS: A prospective, nonblind study was performed in a level I trauma center/community teaching hospital ED. Participants included all nonintubated adult patients with respiratory distress requiring measurement of arterial blood gases (ABGs); 29 patients were enrolled. PetCO2 was measured with a capnography monitor, using both baseline tidal volumes and forced expiratory volumes. The bias between PetCO2 values and simultaneous measurements of PaCO2 by ABG was assessed.
RESULTS: PetCO2, measured with forced expiration, and PaCO2 agreed well, with bias (i.e., average difference) = 0.44 +/- 0.52 kPa (3.3 +/- 3.9 torr). PetCO2 measured with the tidal volume breath produced an unacceptably high bias of 0.82 +/- 0.70 kPa (6.1 +/- 5.2 torr). Levels of agreement between PaCO2 were similar for smokers and nonsmokers and for men and women. The arterial-end-tidal CO2 tension (Pa-etCO2) difference was not related to PaCO2. Pa-etCO2 correlated with age (r = 0.473; p = 0.01), and was significantly higher in patients with pulmonary disease (1.32 +/- 0.56 kPa; 9.9 +/- 4.2 torr) than it was in those without pulmonary disease (0.46 +/- 0.55 kPa; 3.5 +/- 4.1 torr; p < 0.001).
CONCLUSIONS: Noninvasive PetCO2 monitoring may adequately predict PaCO2 in nonintubated ED patients with respiratory distress who are able to produce a forced expiration. PetCO2 is less accurate for PaCO2 with tidal volume breathing and in patients with pulmonary disease.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app