Effect of silicone sound speed and intraocular lens thickness on pseudophakic axial length corrections

S Yang, A Lang, H Makker, E Zaleski
Journal of Cataract and Refractive Surgery 1995, 21 (4): 442-6
8523291
We compared the sound speeds of SLM-1/UV (990 M/sec at 35 degrees Celsius [degrees C]), SLM-2/UV (1090 M/sec at 35 degrees C), and Perspex CQ poly(methyl methacrylate)(2658 M/sec at 35 degrees C). Methods are presented to determine the correction of axial length (CAL) factors for axial length measurements made on pseudophakes with AMO PhacoFlex Si-18, Si-26 (SLM-1/UV), and AMO PhacoFlex II SI-20 or SI-30 (SLM-2/UV) silicone intraocular lenses (IOLs). A CAL is required to avoid potential errors with secondary IOL power predictions; CALs, which are strongly dependent on the material sound speed and less so on lens thickness, ranged from -0.65 mm to -1.20 mm for SLM-1/UV silicone IOLs and from -0.35 mm to -0.55 mm for SLM-2/UV silicone IOLs. The sound speeds of IOL materials varied insignificantly between ambient room temperature (23 degrees C) and intraocular temperature (35 degrees C).

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
8523291
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"

We want to hear from doctors like you!

Take a second to answer a survey question.