Effects of disposable or interchangeable positive end-expiratory pressure valves on work of breathing during the application of continuous positive airway pressure.
Critical Care Medicine 1994 August
OBJECTIVE: To determine which of a series of disposable or interchangeable positive end-expiratory pressure (PEEP) devices functions with the least imposition of inspiratory and expiratory work during continuous positive airway pressure.
DESIGN: Prospective laboratory evaluation performed on a lung model.
SETTING: Research laboratory at a university medical center.
INTERVENTIONS: A spontaneously breathing lung model, created from a training test lung and a volume ventilator, were used to simulate a patient spontaneously breathing at a tidal volume of 0.4 L, peak inspiratory flow of 40 L/min, an inspiration/expiration ratio of 1:2, and a respiratory rate of 20 breaths/min. Ten PEEP valves attached to a continuous high-flow system were evaluated.
MEASUREMENTS AND MAIN RESULTS: All of the PEEP valves studied imposed high levels of both inspiratory and expiratory work of breathing. The BE-171 and BE-142 valves (Instrumentation Industries) imposed the least amount of inspiratory work. In general, imposed inspiratory work accounted for approximately 70% to 80% of total imposed work of breathing.
CONCLUSIONS: All of the disposable/interchangeable PEEP valves that were studied imposed a considerable amount of both inspiratory and expiratory work, even when the continuous flow provided exceeded the peak inspiratory flow demands of the lung model. The primary reason for the high imposed work levels is the high gas-flow resistance of all of the valves studied.
DESIGN: Prospective laboratory evaluation performed on a lung model.
SETTING: Research laboratory at a university medical center.
INTERVENTIONS: A spontaneously breathing lung model, created from a training test lung and a volume ventilator, were used to simulate a patient spontaneously breathing at a tidal volume of 0.4 L, peak inspiratory flow of 40 L/min, an inspiration/expiration ratio of 1:2, and a respiratory rate of 20 breaths/min. Ten PEEP valves attached to a continuous high-flow system were evaluated.
MEASUREMENTS AND MAIN RESULTS: All of the PEEP valves studied imposed high levels of both inspiratory and expiratory work of breathing. The BE-171 and BE-142 valves (Instrumentation Industries) imposed the least amount of inspiratory work. In general, imposed inspiratory work accounted for approximately 70% to 80% of total imposed work of breathing.
CONCLUSIONS: All of the disposable/interchangeable PEEP valves that were studied imposed a considerable amount of both inspiratory and expiratory work, even when the continuous flow provided exceeded the peak inspiratory flow demands of the lung model. The primary reason for the high imposed work levels is the high gas-flow resistance of all of the valves studied.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app