CLINICAL TRIAL
COMPARATIVE STUDY
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
Add like
Add dislike
Add to saved papers

Mechanical ventilators optimized for pediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation.

OBJECTIVES: a) To investigate whether the patient work of breathing needed to trigger inspiration is affected by the type of ventilator delivering pressure-support ventilation for mechanically ventilated pediatric patients. b) To determine whether changes in oxygen consumption (VO2) trend with changes in work of breathing and would thus be helpful in tracking work of breathing.

DESIGN: Prospective study.

SETTING: Pediatric intensive care unit at a university hospital.

PATIENTS: Nine mechanically ventilated patients (2 to 75 months of age).

INTERVENTIONS: While maintaining a constant pressure-support ventilation level, patients were alternately supported with the Siemens Servo 900C, the Bird VIP, and the Newport Wave E200 ventilators in random order.

MEASUREMENTS AND MAIN RESULTS: Work of breathing, defined as the integral of the pressure-volume curve corresponding to negative pressure, was calculated with a pulmonary monitoring system. VO2 was measured with a metabolic cart. Patient distress levels were assessed using the COMFORT scale, a behavioral scoring system. Mean values (20 breaths/patient) for measured variables with each ventilator were compared using analysis of variance and Scheffé tests, with p < .05 indicating statistical significance. The lowest VO2 (103 +/- 35 mL/min/m2) and work of breathing (24 +/- 15 g.cm/m2) were achieved with the Bird VIP ventilator and were significantly (p < .05) lower than those values obtained with either the Siemens Servo 900C (VO2 147 +/- 33 mL/min/m2; work of breathing 49 +/- 18 g.cm/m2) or the Newport Wave E200 (VO2 122 +/- 33 mL/min/m2; work of breathing 35 +/- 15 g.cm/m2). Also, the values of work of breathing and VO2 obtained using the Newport Wave E200 were significantly (p < .05) lower than those values obtained using the Servo 900C. No change in behavioral distress occurred when the ventilators were changed. In all patients, there was a clear similarity in the trends of VO2 and work of breathing.

CONCLUSIONS: We conclude that VO2 and work of breathing may be reduced significantly using the latest generation of mechanical ventilators optimized for infant and pediatric use. Because work of breathing is less with the Bird VIP than the other two ventilators tested, leading to a corresponding decrease in VO2, we suggest that the Bird VIP better adapts the patient to the ventilator and may facilitate weaning from ventilatory support. We also suggest that changes in VO2 might be helpful in tracking changes in work of breathing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app