Transluminal placement of endovascular stent-grafts for the treatment of descending thoracic aortic aneurysms.
New England Journal of Medicine 1994 December 30
BACKGROUND: The usual treatment for thoracic aortic aneurysms is surgical replacement with a prosthetic graft, but the associated morbidity and mortality are considerable. We studied the use of transluminally placed endovascular stent-graft devices as an alternative to surgical repair.
METHODS: We evaluated the feasibility, safety, and effectiveness of transluminally placed stent-graft to treat descending thoracic aortic aneurysms in 13 patients over a 24-month period. Atherosclerotic, anastomotic, and post-traumatic true or false aneurysms and aortic dissections were treated. The mean diameter of the aneurysms was 6.1 cm (range, 5 to 8). The endovascular stent-grafts were custom-designed for each patient and were constructed of self-expanding stainless-steel stents covered with woven Dacron grafts.
RESULTS: Endovascular placement of the stent-graft prosthesis was successful in all patients. There was complete thrombosis of the thoracic aortic aneurysm surrounding the stent-graft in 12 patients, and partial thrombosis in 1. Two patients initially had small, residual patent proximal tracts into the aneurysm sac, but both tracts thrombosed within two months after the procedure. In four patients, two prostheses were required to bridge the aneurysm adequately. There have been no deaths or instances of paraplegia, stroke, distal embolization, or infection during an average follow-up of 11.6 months. One patient with an extensive chronic aortic dissection required open surgical graft replacement four months later because of progressive dilatation of the arch.
CONCLUSIONS: These preliminary results demonstrate that endovascular stent-graft repair is safe in highly selected patients with descending thoracic aortic aneurysms. This new method of treatment will, however, require careful long-term evaluation.
METHODS: We evaluated the feasibility, safety, and effectiveness of transluminally placed stent-graft to treat descending thoracic aortic aneurysms in 13 patients over a 24-month period. Atherosclerotic, anastomotic, and post-traumatic true or false aneurysms and aortic dissections were treated. The mean diameter of the aneurysms was 6.1 cm (range, 5 to 8). The endovascular stent-grafts were custom-designed for each patient and were constructed of self-expanding stainless-steel stents covered with woven Dacron grafts.
RESULTS: Endovascular placement of the stent-graft prosthesis was successful in all patients. There was complete thrombosis of the thoracic aortic aneurysm surrounding the stent-graft in 12 patients, and partial thrombosis in 1. Two patients initially had small, residual patent proximal tracts into the aneurysm sac, but both tracts thrombosed within two months after the procedure. In four patients, two prostheses were required to bridge the aneurysm adequately. There have been no deaths or instances of paraplegia, stroke, distal embolization, or infection during an average follow-up of 11.6 months. One patient with an extensive chronic aortic dissection required open surgical graft replacement four months later because of progressive dilatation of the arch.
CONCLUSIONS: These preliminary results demonstrate that endovascular stent-graft repair is safe in highly selected patients with descending thoracic aortic aneurysms. This new method of treatment will, however, require careful long-term evaluation.
Full text links
Trending Papers
2023 ACC/AHA/ACCP/HRS Guideline for the Diagnosis and Management of Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines.Circulation 2023 November 31
ANCA-associated vasculitis - Treatment Standard.Nephrology, Dialysis, Transplantation 2023 November 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app