We have located links that may give you full text access.
A new wireless device for bedside assessment of tip location of central venous access devices using intracavitary ECG: A retrospective review.
Journal of Vascular Access 2024 August 22
Intracavitary electrocardiography (IC-ECG) is a non-invasive method for intraprocedural tip location during central venous catheterization. Over the last 20 years, the IC-ECG method has been thoroughly investigated and many studies have confirmed its wide applicability and feasibility, as well as its great accuracy and safety in different populations of patients and in different types of central venous access devices (CVADs). This retrospective study presents a two-center experience with tip location of central venous catheters using IC-ECG, by means of a new wireless portable device. Tip location was performed during 983 insertions of CVADs at bedside, using this new device. Clinical conditions with limited or no applicability of IC-ECG were excluded. The feasibility of IC-ECG, as performed using the novel device, was 92.2%. This wireless device may play an important role in central venous catheterization at bedside, since it is pocket-sized, particularly easy to use, and easy to sanify after use.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app